High and Low Temperature and Humidity Chamber 150L IEC60068-2-78
 
 
Complied Standards & Clauses


 
IEC60068-2-1:2007 “Environmental Testing-Part 2-1: Tests-Test A: Cold, IDT”
IEC60068-2-30:2005 “Environmental Testing-Part 2-30: Tests-Test Db: Damp heat, cyclic (12h+12h cycle), IDT”
IEC60068-2-78:2012 “Environmental Testing-Part 2-78: Tests-Test Cab: Damp heat, steady state, IDT”
IEC60068-2-2:2007 “Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT”
 
Equipment Overview


 
The purpose is to confirm the product's reliability of storage and transportation under high and low temperature alternating heat and humidity. This test chamber is used for cold, heat, humidity and dryness testing of raw materials and devices such as electronics, electricians, electrical appliances, plastics, etc. 

 
Technical Parameters


 

Technical Parameters
Control & OperationPLC and touch screen intelligent control system

Chamber Size
 

Inner SizeExternal Size
500*600*500 mm780*1650*1550 mm
Temperature Range-70 °C ~+150°C
Humidity Range20%~98%RH
Temperature fluctuation±0.5℃
Humidity fluctuation≤2.5%
Temperature deviation≤±2.0℃
Humidity Deviation≤±3%
Heating Rate≧2-3°C/min
Cooling Rate≧1°C/min, average no load
 
c) Controller specifications

Technical indicators:
1. 7-inch color touch screen
2. Two control modes: program/fixed value
3. Sensor type: two PT100 inputs (electronic sensor input optional)
4. Output mode: voltage pulse (SSR)/control output: 2 (temperature/humidity)/2 4-20mA analog output/16 relay outputs (passive)
5. Control signal: 8 IS control signals/8 T control signals/4 AL control signals
6. Alarm signal: 16 DI external obstacle alarms
7. Temperature measurement range: -90.00℃--200.00℃, (optional -90.00℃--300.00℃) error ±0.2℃
8. Humidity measurement range: 1.0%--100%RH, error ±1%RH
9. Communication interface: (RS232/RS485, maximum communication distance 1.2km [optical fiber up to 30km])
10. Interface language type: Chinese/English
11. Equipped with English input function
12. Can be used with a printer (USB function optional)
13. A variety of signal combination relay outputs, and the signals can be subjected to logical operations (NOT, AND, OR, NOR, XOR), referred to as PLC programming ability
14. Diversified relay control methods: parameter->relay mode, relay->parameter mode, logic combination mode, composite signal mode
15. Program editing: 120 groups of programs can be programmed, and each group of programs can program up to 100 segments

Controller technical parameters

Accuracy: temperature ±0.1℃+1digit, humidity ±1%R.H+1digit
Resolution: temperature ±0.01℃, humidity ±0.1%R.H.
Temperature slope: 0.1~9.9 can be set.
It has upper and lower limit standby and alarm functions.
Temperature and humidity input signal dry and wet bulb PT100x2.
9 sets of P.I.D control parameter settings, P.I.D automatic calculation.
Dry and wet bulb with automatic correction screen function

 
 
Details

 

 

 

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Sinuo Testing Equipment Co. , Limited
Location Room 101, 1st Floor, No. 6, Third Street, Pingshan Industrial Zone, Shibi Street, Panyu District, Guangzhou, China
Contact Person Zoe Zou

Request A Quote

Please check your email address.
Your message must be at least 20 characters.