MOX20 MOX-20 AAE29-210 Gas Oxygen Sensor Use Anesthesia Machines Oxygen Concentrators
Price:
Negotiable
MOQ:
1
Delivery Time:
3-5work days
Product Description
MOX-20 Gas oxygen sensor
Basic Parameters:
- Measurement Range: 0%–100% O₂
- Output Signal: 13–16 mV in air
- Circuit Interface: 3-pin (Molex)
- Accuracy & Repeatability: <1% O₂ by volume (after calibration in 100% O₂)
- Linearity Error: <3% relative error
- Response Time (T90): <12 seconds (to 90% of final value)
- Zero Drift: <200 μV after 5 minutes in 100% NH₃
- Temperature Influence: -0.03%/°C relative to 25°C
- Pressure Influence: Proportional to O₂ partial pressure
- Mechanical Shock Resistance: <1% relative error after 1m drop
- Operating Temperature: 0°C–50°C
- Temperature Compensation:
- Built-in thermistor compensation
- ±3% relative error at 25°C–40°C
- ±8% relative error (steady-state) at 0°C–50°C
- Operating Humidity: 0%–99% RH (tolerates condensation for hours)
- Storage Temperature: -20°C–50°C (recommended: 5°C–15°C)
- Recommended Load Resistance: ≥10 Ω
- Warm-up Time: <30 minutes after sensor replacement
- Rated Service Life: ≥1,000,000% O₂·h (volume-oxygen hours)
- Weight: ~28 g
Applications:
The MOX-20 oxygen sensor is primarily used in medical devices such as:
- Ventilators
- Anesthesia machines
- Oxygen concentrators
- Oxygen analyzers for measuring O₂ concentration in respiratory gas mixtures.
Manufacturer:
For detailed technical specifications and application notes, please refer to the official datasheet.
|
Parameter |
Value |
|---|---|
| Model | MOX-20 |
| Sensor Type | Electrochemical oxygen sensor |
| Measurement Range | 0%–100% O₂ (by volume) |
| Output Signal in Air | 13–16 mV |
| Interface | 3-pin Molex connector |
| Accuracy | <1% O₂ (after calibration in 100% O₂) |
| Linearity Error | <3% relative error across full range |
| Response Time (T90) | <12 seconds (to 90% of final value) |
| Zero Drift in NH₃ | <200 μV after 5 min in 100% NH₃ |
| Temperature Coefficient | -0.03%/°C (relative to 25°C) |
| Pressure Dependence | Proportional to O₂ partial pressure |
| Shock Resistance | <1% error after 1m drop test |
| Operating Temperature | 0°C–50°C |
| Storage Temperature | -20°C–50°C (recommended: 5°C–15°C) |
| Humidity Range | 0%–99% RH (non-condensing, tolerates short-term condensation) |
| Temperature Compensation | Built-in thermistor; ±3% error at 25°C–40°C, ±8% at 0°C–50°C |
| Recommended Load Resistance | ≥10 Ω |
| Warm-up Time | <30 minutes after installation |
| Service Life | ≥1,000,000% O₂·h (volume-oxygen hours) |
| Weight | ~28 g |