Kidney Function Calibrator NGAL β2-MG Cys-C RBP H-ALB Reagent Item Six-points Calibrator Matching Reagent Analyzer

 

Testing method:
1. Treat the calibrator as the sample to be tested, and place it in the sample loading area of the instrument after mixing. Note that when mixing Avoid air bubbles.

2. During the detection, the system detects the luminescence value (RLU value) of each calibration point, and calculates the luminescence value according to the nominal concentration and luminescence value.
The calibration curve is automatically fitted by the instrument. In order to ensure the reliability of the calibration curve, the determination of the calibrator should be make a double.

 

Storage and expiration date:
Stored at 2-8°C, the product is valid for 18 months.

Reagents should be used within 1 month after opening, and stored at 2-8°C away from direct sunlight.

 

CLIA/POCT Calibrator of Kidney Function:

Kidney Function Item NGAL
β2-MG
Cys-C
RBP
H-ALB
Six-point Calibrator Calibrator A
Calibrator B
Calibrator C
Calibrator D
Calibrator E
Calibrator F
Calibrator Performance Appearance Each component should be complete, the internal and external packagipg should be complete, the label should be clear, and there should be no leakage of liquid reagent.
Volume The volume of each component shall be accurately weighed, and it shall not be lower than the nominal volume.
Accuracy The relative deviation should be within ± 10%.
Homogeneity The homogeneity between bottles CV ≤ 10%.
Stability Under the storage condition of 2 ~ 8 , the validity period of IL-6 calibrator is 12 months. Take the samples stored within 1 month after the validity period to test the appearance, volume, accuracy and uniformity index, which should meet the standard requirements.

 

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Taizhou Zecen Biotech Co.,Ltd.
Location Building G62, Phase 4 Plant, China Medical City, Taizhou City, Jiangsu Province
Contact Person Liu

Request A Quote

Please check your email address.
Your message must be at least 20 characters.