Sign In | Join Free | My bjjsgy.com |
|
Categories | Rubber Testing Machine |
---|---|
Brand Name: | ZL |
Model Number: | ZL-3001 |
Certification: | CE, ISO |
Place of Origin: | China |
MOQ: | 1 SET |
Price: | Negotiable |
Payment Terms: | T/T, L/C, Western Union |
Supply Ability: | 50 SET PER MONTH |
Delivery Time: | 7-10 Days |
Packaging Details: | Packaged in wooden case (fumigation free) |
Power: | Electronic |
Temperature resolution: | 0.1°C |
Motor Roating speed: | 100 times/min |
Torque Resolution: | 100 times/min |
Air compression: | 0.5Mpa~0.65Mpa |
Warranty: | 12 month |
Weight: | 250kg |
Torque range: | 0-5N.M,0-10N.M,0-20N.M |
Company Info. |
Dongguan Zhongli Instrument Technology Co., Ltd. |
Verified Supplier |
View Contact Details |
Product List |
Testing Rubber Industry Used No Rotor Rheometer
The main technical parameters:
Model | Without Rotor Rheometer Rubber Testing Machine |
Temperature tester | Room temperature to 200°C |
Temperature accuracy | ≤ ±0.3°C |
Temperature resolution | 0.1°C |
Torque range | 0-5N.M,0-10N.M,0-20N.M |
Torque resolution: | 0.001NM |
Motor rotating speed | 100 times / min |
Rotor range | Adopt 4 setting: 25, 50, 100, 200lbs per inch |
Die cavity frequency | 1.7Hz |
Sway angle | ±0.5°, ±1°, ±3° |
Air compression | 0.5Mpa--0.65MPa |
Ambient temperature | 0-35°C |
Test result | Printing:date,time,temperature,vulcanization curve,temperature curve,ML,MH,ts1,ts2,t10,t50,Vc1,Vc2. |
Dimension | 131*56*62cm |
Power | Single phase, 220V±10%, 50Hz |
Display | Windows 7 + software |
Weight | 250kg |
Summary:
The use of single-chip control, including: host, temperature
measurement, temperature control, data acquisition and processing,
sensors and electrical interlocking and other components. The
measurement and temperature control circuit consists of a
temperature controller, a platinum resistor, and a heater. It can
automatically track changes in the power grid and ambient
temperature, automatically correct PID parameters, and achieve
rapid and accurate temperature control. The data acquisition system
and the electromechanical chain complete the automatic detection of
the torque signal of the rubber vulcanization process, automatic
real-time display of temperature and set values. After
vulcanization is completed, automatic processing, automatic
calculation, printing vulcanization curve and process parameters.
Show vulcanization time, sulfurizing power torch.
Corresponding standard:
Accord with GB/T16584”rubber- no rotor rheometer to measure the characteristics of vulcanized”ISO6502:1991 and ASTMD5289-95
Download Brief:
ZL-3001 No Rotor Rheometer.pdf
The No Rotor Rheometer is the most widely used instrument in the rubber processing industry to control rubber quality, rapid inspection and basic rubber research. It provides accurate data for the optimal rubber formula combination, and can accurately measure the scorch time, positive vulcanizer time, and vulcanization. Index and parameters such as maximum and minimum torque.
![]() |