Diagnostic Kit for Cystain C Rapid Test Kit IFA Colloidal Gold IVD Blood Diagnostic
【Product name】
Carcinoembryonic Antigen (CEA) Rapid Quantitative Test(Fluorescence immunoassay)
【Intended use】
This kit is used for quantitative determination of CEA in human whole blood, plasma and serum.
Carcinoembryonic antigen (CEA) is a tumor associated antigen, first described in 1965 by Gold and Freedman1. It is a cell-surface glycoprotein with a molecular weight of 180-200kD, that occurs in high levels in colon epithelial cells during embryonic development. Levels of CEA are significantly lower in colon tissue of adults, but can become elevated when inflammation or tumours arise in any endodermal tissue, including the gastrointestinal tract, respiratory tract, pancreas and breast.
An over expression of CEA protein has been detected in a variety of adenocarcinomas, including gastric, pancreatic, small intestine, colon, rectal, ovarian, breast, cervical and non-small-cell lung cancers. CEA is also expressed by epithelial cells in several non-malignant disorders, including diverticulitis, pancreatitis, inflammatory bowel disease, cirrhosis, hepatitis, bronchitis and renal failure and also in heavy smokers.
Therefore CEA should not be regarded as a tumour-specific marker for the screening of a general population for undetected cancers. However, the determination of CEA levels provides important information about patient prognosis, recurrence of tumours after surgical removal and effectiveness of therapy.
【Components】
| Name | Quantity | Component |
| Test cards | 25 | It is composed of fluorescent pad (coated with fluorescent labeled CEA monoclonal mouse antibody), nitrocellulose membrane (coated with CEA monoclonal mouse antibody and Goat anti mouse IgG antibody), absorbent paper and backing |
| Sample diluent | 25(300μL/tube) | Phosphate buffer |
| ID card | 1 | With specific stand curve file |
The components in different batches of kits cannot be used interchangeably.
【Storage conditions and validity】
The kit should be stored at 4℃~30℃, out of direct sunlight. It is valid for 18 months. The test card should be used within 15 minutes after unsealing under the environment of 15℃~30℃ and 20% ~ 90% relative humidity.
The production date, batch number and expiration date are shown in the outer package of the product.
【Applicable instruments】
NIR-1000 dry fluoroimmunoassay analyser produced by WWHS Biotech. Inc.
【Procedure】
- Before the test, please read the instructions completely. If the test card and sample are stored in cold storage, they should be balanced at room temperature (15-30)℃ for not less than 30min before use.
- Start NIR-1000 dry fluoroimmunoassay analyser according to the instruction manual of the instrument, and carry out quality control verification according to the instruction manual of the instrument(Note: the reagent has been calibrated in advance, and the calibration curve parameters of each batch of reagent have been stored in the information card. The information card is inserted before use, so it is not necessary to calibrate again, and the test can be carried out only after the quality control is passed. Otherwise, the cause should be found out before testing.)
- Remove the test card from the aluminum foil bag and use it within 15 minutes.
- Place the test card on a clean horizontal table and mark it horizontally.
- Mix 100µL of patient sample with 300µL of sample diluent. Apply 100µL of diluted samples to the well of the test card.
- Insert the test card into NIR-1000 dry fluoroimmunoassay analyser, read and record the results at 10 minutes after addition of samples, then dispose of used test appropriately.
【Limitations of methods】
- This kit is only used to detect human plasma/whole blood samples
- Due to the limitations of immunoassay methods of antigen and antibody reaction, the results cannot be used as the only basis for clinical diagnosis, but should be evaluated with all the existing clinical and experimental data.
- The content of triglyceride in the sample shall not exceed 15mg/ml, the content of hemoglobin shall not exceed 5mg/ml, and the content of bilirubin shall not exceed 0.5mg/ml, and the relative deviation of the test results shall not exceed ±15%.
- When the concentration of CEA in the sample is less than 20000ng/ml, there is no hook effect.
- HAMA effect was not produced when the concentration of human anti rat in the sample was less than 50ng/ml.
- When RF concentration in the sample is less than 2000IU/ml, the relative deviation of the test results is within ±15%.
【Performance】
1. Limits of detection
No more than 1ng /ml.
2. Accuracy
The relative deviation from the target value is within ±15%.
3. Precision
The within and between assay coefficient of variations are within 15%.
4. Linear range
Within the linear range (1 ~ 500ng/ml), the linear correlation coefficient R≥0.990.