3 5 12 Lead Temperature Respiration ECG Sensor Module IEC601-1
OEM 3/5/12 Lead Temperature Respiration ECG Module for Patient Monitor
| Adopting digital filter technology |
| Selectable 50Hz or 60Hz band stop filter |
| 500Hz,7 /12 lead synchronous sample and analysis |
| Character Classification Up to twelve sort, and 3-channel ST Segment Monitoring |
| Anti high-frequency electric-knife |
| Adopting single 5V power supply,low consumption and safely insulation |
| 5V TTL Communication |
| 2-channel temperature detect ,and two sort of sensor model |
| Detecting respiration by impedance between RA-LL |
|
8-channel ECG waveform |
|
ST Segment Analysis |
|
Heat rate calculation |
|
Arrhythmia analysis |
|
Lead off status indication |
|
Lead mode |
12 leads (R, L, F, N, C1~C6 or RA, LA, LL, RL, V1~V6 ) |
|
Selection |
I, II, III, aVR, aVL, aVF, Vx, CAL |
|
Gain |
´0.25V/mV, ´0.5V/mV, ´1V/mV, ´2V/mV |
|
Heart rate(HR) |
Range: Adult 15 ~ 300 bpm Neo/Ped 15 ~ 350 bpm Accuracy: ±1% or ±1bpm Sensitivity: > 200 uVpp |
|
Differential Input Impedance |
> 5 MΩ |
|
CMRR |
> 100 dB |
|
Electrode offset potential |
±400mV |
|
Insulation Intensity |
4000VACrms |
|
Baseline Recovery |
< 3 s, after Defi. |
|
ECG Signal Range |
>±5 mVpp |
|
Bandwidth |
Diagnostic 0.05 ~ 100 Hz Monitor 0.5 ~ 40 Hz Surgery 1 ~ 20 Hz Strong filter 5 ~ 20Hz |
|
Calibration Signal |
1mVpp,Accuracy ±3% |
|
ST segment monitoring range |
-2.0mV ~2.0mV |
|
Arrhythmia analysis |
ASY,VF/VTA,CPT,RUN,BGM,TGM,R on T(ROT),VPB,TAC,BRD,MIS,PNC,PNP |
|
Waveform |
1 ch |
| Respiration Rate |
dult 7-120bpm
Ped and Neo 7-150bpm
Resolution 1bpm
Accuracy ±2bpm
|
| Suffocation alarm | 10-40s Gain 10levels |
| Baseline Impedance |
500 - 4000Ω |
| Measuring Range |
0.5 - 5Ω |
|
Channel Number |
2 ch |
|
Probe type |
2K and 10K probe |
|
Range |
0-50°C |
|
Resolution |
0.1°C |
|
Accuracy |
±0.1°C |