VGB600S Stainless Steel Laboratory Glove Box 600x450x420mm with Vacuum System and Siemens PLC Control
The VGB600S Stainless Steel Laboratory Glove Box is a heavy-duty anaerobic chamber designed for controlled environment applications. Featuring a stainless steel construction with airlock chamber, gauges, and vacuum flanges, this glove box operates under vacuum to effectively remove moisture and oxygen without requiring expensive inert gas purging.
- MOCVD technology in semiconductor industry
- Special lamps (HID) development and production
- Microelectronics, laser and plasma welding
- OLED/PLED development and production
- Lithium-ion battery development and production
- Inorganic/Organic Chemistry Research
- Metal halide/metal powder/ceramic powder research
- Nanotechnology research materials
- Catalyst research and development
- Siemens PLC control system with 6" touch screen display
- Closed-loop gas recirculation system, oil-free and vacuum-tight
- Automatic pressure control with +/-12mbar range
- Automatic purging system for gas efficiency and solvent safety
- 9 3/4" butyl gloves included for immediate use
- Foot switch for convenient pressure regulation
- Automatic data logging system
- One year limited warranty with lifetime support
| Component | Specifications |
|---|---|
| Glove Box Chamber | Dimensions: 600mm(W) × 450mm(D) × 420mm(H) Material: Stainless SUS304, 3mm thickness Max Positive Pressure: 0.05 MPa Max Vacuum Level: -0.05 MPa |
| Antechamber | Dimension: ø210mm, length 260mm Material: Stainless SUS304 Max Positive Pressure: 0.1 MPa Max Vacuum Level: -0.1 MPa |
| Front Panel Window | Dimension: 500mm(W) × 165mm(H) Removable tempered window with 2 glove ports (12mm) |
| Gloves | 1 pair of Latex gloves, 190mm length |
| Lighting System | 220V, 50Hz |
| Connectors | Standard KF-25, KF-40 connectors and needle valves |
- Oxygen monitoring unit (0-1000 PPM)
- Moisture monitoring unit (0-1000 PPM)
- Automatic Pressure Controller
- Automatic Humidity Controller (1-99% RH)
- Solvent removal system
- Open loop/Closed-loop HEPA Filters
- High speed vacuum pump (4L/S)
- Gas Purification System
Important Notes: Always check joints, gloves, sealing elements, and pipelines for air-tightness before and during operation. Corrosive gases are prohibited as they will damage sensors. The air-lock chamber inlet and outlet must not be opened simultaneously to prevent external gas contamination. By closing vacuum flanges, the glove box can be vacuumed by mechanical pump up to 0.5 torr before filling inert gas, achieving moisture below 180 ppm and oxygen below 200 ppm while saving 1-2 tanks of inert gas per operation.
| Item | Description | Quantity |
|---|---|---|
| 1 | VGB-600S SS Glove box | 1 unit |
| 2 | Power cable | 1 pair |
| 3 | Gloves (spare use) | 2 pcs |
| 4 | Vacuum rubber hose | 1 pcs |
| 5 | Sealing gasket of glove port (spare use) | 3 pcs |
| 6 | Sealing gasket of glove box door (spare use) | 2 pcs |
| 7 | Vacuum grease | 1 box |
| 8 | Glove clamps | 3 pcs |
| 9 | Hose clamps | 1 pack |
| 10 | Polyurethane tube | 1 pc |
| 11 | User's manual CD | 1 pc |