Guangdong Sanwood Technology Co.,Ltd
                                                                                                           
Verified Supplier
31 Years
Since 1995
Menu

EN14683 Respirator BFE Tester For Medical Mask Test Meet

Price Negotiable
Price: Negotiable
MOQ: 1
Delivery Time: 1 week
Brand: SANWOOD
Product Description

Respirator Bacterial Filtration Efficiency (BFE) Tester (required item for Medical face mask)

 

 

Application:

Applied to performance test of the bacterial filtration efficiency of surgical masks by the metrological verification department, scientific research institutes, mask manufacturers and other relevant departments.

 

Standards:

YY0469-2011 < Medical grade surgical face mask> Fig. B.1 and Fig. B.2 in Appendix B

 

Features:

1. Negative pressure test system to ensure the safety of operators;

2. Built - in peristaltic pump in the negative pressure cabinet, A, B, four Andersen;

3. The peristaltic pump flow can be set;

4. The flow rate of night spray of special microorganism aerosol generator can be set;

5. Good atomize effect;

6. Embedded high-speed industrial microcomputer control;

7. 10.4 inch industrial high brightness color touch screen;

8. USB interface, support U disk data storage;

9. Cabinet built-in high brightness lights;

10. Front opening and closing glass door, convenient for experimenters to observe and operate;

11. In stock 

 

 

Specifications:

Main Specifications Parameter range Resolution Accuracy
A Sampling Flow 28.3L/min 0.1L/min ±2.5%
B Sampling Flow 28.3L/min 0.1L/min ±2.5%
Spray Flow (8~10)L/min 0.1L/min ±2.5%
Peristatic Pump Flow (0.006~3.0)ml/min 0.001ml/min ±2.5%
A Road Flow Meter (-20~0)KPa 0.01KPa ±2.5%
B Road Flow Meter (-20~0)KPa 0.01KPa ±2.5%
Pressure Before Spray Flowmeter (0~300)KPa 0.1KPa ±2.5%
Environmental Temperature (-40~99)℃ 0.1℃ ±2.5%
Negative Pressure In Aerosol Chamber (-90~-120)Pa 0.1Pa ±2.0%
Negative Pressure Of Cabinet -50~-200Pa
Data Storage Capacity >100000 Groups
Vortex Mixer Test Tube Specification and Quantity Φ16×150 mm test tubes, Eight
Characteristics of High Efficiency Air Filter The filtration efficiency of particles above 0.3 um ≥99.99%
Median Mass Diameter Of Aerosol Generator 3.0± 0.3μm,Geometric standard deviation ≤1.5
Double Route 4 Andersen

Level I >7μm,Ⅱ4.7-7μm,

Level Ⅲ 3.3-4.7μm,Level Ⅳ 2.1-3.3μm,

Level Ⅴ 1.1-2.1μm,Level Ⅵ 0.6-1.1μm

Aerosol Chamber Size(L*D*T) 600*85*30mm
Total Particles of Positive Quality Control Sampler 2200±500cfu
Ventilation flow of negative pressure cabinet ≥5m3/min
Negative Cabinet Door Size(W*D) 1000×730mm
Mainframe Size(W*D*H) 1180×650×1300mm
Support Dimensions(W*D*H) 1180×650×600mm,Adjustable height within 10 cm
Instrument Noise <65dB(A)
Weight About 150kg
Power Supply AC220V±10%,50Hz
Power Dissipation <1500W

 

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Guangdong Sanwood Technology Co.,Ltd
Contact Person Nancy Shen

Request A Quote

Please check your email address.
Your message must be at least 20 characters.