ICU Ventilator Shangrila590
Shangrila590
- The first CE marked ICU ventilator in China
- Pneumatically driven and electronically controlled for pediatric and adult patient
- Detachable 10.4" TFT color screen decreases the possibility of cross-infection
- Easy operation by using navigator wheel knob & touch key
- Three-Ievel light and sound alarm with short text description
- Sensitive pressure trigger and flow trigger with base flow
- Lung mechanics monitoring and loop function
- Emergency start function
- Synchronized nebulizer
- Durable built-in inspiration and expiration flow sensor, protected from water condensation
- Built-in battery
Technical Specification
Ventilation Mode
• Volume Controlled (VCV), Assist/Control
• Pressure Controlled (PCV), Assist/Control
• Synchronized Intermittent Mandatory Ventilation (SIMV)
• Volume-controlled breaths (V-SIMV)
• Pressure-controlled breaths (P-SIMV)
• Pressure Support (PSV)
• Spontaneous ventilation (SPONT)
• Continuous Positive Airway Pressure (CPAP)
• Combined settings: VCV + SIGH, SIMV (VCV) + PSV,SIMV (PCV) + PSV
Sigh
Parameter
• Tidal volume: 50~1500
• Respiratory rate: 4~100bpm (VCV & PCV)ml,1~40bpm (V-SIMV & P-SIMV)
• Inspiration time: 0.1~12s
• Inspiratory pause time: 0~4s
• FiO2: 21%~100%
• Trigger sensitivity: Pressure (-2~0kPa, above PEEP),Flow (0.5~20LPM, with base flow)
• PEEP: 0~4kPa
• Psupport: 0~6kPa
• Pcontrol: 0.5~6kPa
• Nebulizer: 30 minutes
• Emergency start: Yes
• Key lock: Yes
Monitoring
• Pressure values: Paw, Pplat, Pmean, PEEP
• Volume/Flow values: VTI, VTe, MV, MVspont
• Time values: ftotal, fspont, I:E
• Oxygen monitoring: without O2 sensor
• Real time curves: Pressure-Time, Flow-Time, Volume-Time,Pressure-Volume loop, Flow-Volume loop
• Respiratory mechanics: dynamic and static compliance,resistance, auto PEEP
Alarm
• MV high, MV low, Circuit disconnected, Paw high, Paw low,
• Airway pressure continue high, VTe high, f spont high,
• Tapnea, FiO2 low, FiO2 high Power supply failure,
• Nebulizer on, Battery low, Battery exhausted
• Air supply down, O2 supply down, Air & O2 supply down
Technical Data
• Screen: 10.4”TFT color screen (detachable)
• Gas supply: O2, Air (AlI gas must be medicallevel), 0.28~0.6MPa
• Power supply: AC110~240V, 50Hz/60Hz, 65VA
• Maximum security pressure: 8kPa
• Compliance: 4mL/100Pa
• Noise: ≤65dB(A)
• Communication interfaces: RS232 port, VGA port
• Dimensions (HxWxD): 400x303x250 mm
• Weight: 15kg
Environment Requirement
• Temperature: 5˫_40˫(Operation)20˫_55˫(Operation)
• Relative humidity: ≤90%, non-condensing (Operation),≤93%, non-condensing (Storage)
• Atmospheric pressure: 86~106kPa (Operation),86~106kPa (Storage),500~800mmHg / 3565~-440m (Operation),375~800mmHg / 5860~-440m (Storage)
Get in Touch