ASTM D93 Fully Automatic Oil Analysis Equipment Closed cup Flash Point Tester

Price Negotiable
Price: negotiated
MOQ: 1 Unit
Delivery Time: Negotiable
Brand: SKYLINE
Product Description

ASTM D93 Fully Automatic Oil Analysis Equipment Closed cup Flash Point Tester


 
Main Features Of Closed cup Flash Point Tester
1.It adopts LCD screen to display. Full English man-machine dialog interface. It can preset the parameters of expected flash point, sample mark number, atmospheric pressure, test date, etc. It has menu to prompt and input function of guide type.
2.It adopts simulation tracking display the function curve of temperature rising and test time. It has prompt functions of English mis-operation, test date and test time.
3.It is equipped with standard RS-323,485 computer port. The lower computer can save 100 groups of historical data. The test data can be saved for a long time, transmitted and modified if the instrument being connected with computer.
4.It can correct the influence of atmospheric pressure and calculate the correction automatically.
5. Differential detection. Automatically correct the systematic deviation.
6. Automatically open the lid, ignite, detect and print test data. The test arm lifting up and down automatically.
7. It adopts electric ignition to fire light the gas flame. It will light only need to press the button.
8. Humanized design, beautiful and safe. Easy to operate.

 

Product parameters
1 .Power supply:AC(220±10%)V,50Hz
2. Temperature measurement Range: room temperature~300°C
Repeatability:≤3°C
Reproducibility:≤6°C
Accuracy:0.1°C
3. Heating rate: Procedure A:(5~6)°C/mi
Procedure B:(1~1.6)°C/min
Automatic control and manually adjustable.
4.Stirring rate
(1) Procedure A: (90~120)RPM
(2) Procedure B: (250±10)RPM
Automatic control and manually adjustable.
5.Igniting mode: Electric ignition. Gas flame diameter:3.2mm~4.8mm
6.Working condition: ambient temperature:(10~40)°C
relative humidity:≤80%
7. Maximum power consumption 500W
8. Dimension 520mm×360mm×310mm

1 Power supply AC(220±10%)V, 50Hz.
2 Flash point determination Range: Ambient to 300℃.
Meet requirement of GB/T 261-2008
Meet requirement of GB/T 261-2008
Resulotion: 0.1℃
3 Heating rate Procedure A: (5~6)℃/min
Procedure B: (1~1.5)℃/min
4 Stirring rate Procedure A: (90~120)RPM,
Procedure B: (250±10)RPM
Automatic control and manually adjustable.
5 Igniting mode
Igniting source: gas(or other civilian fuels)Electric ignition.
Gas flame diameter: 3.2mm~4.8mm
6 Ambient temperature ≤ 40℃
7 Relative humidity ≤ 80%
8 Maximum power consumption 500W
9 Overall dimension 410mm×360mm×310mm

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company SKYLINE INSTRUMENTS CO.,LTD
Location No.1802 Room 6# Building Universal Commerce Center Nancheng District Dongguan Guangdong China
Contact Person Neo

Request A Quote

Please check your email address.
Your message must be at least 20 characters.