Medical equipment Mainstream Nihon Kohden EtCO2 Sensor
product description
|
Item
|
Value
|
|
Vailable Brand
|
Nihon Kohden
|
|
MOQ
|
1PCS
|
|
Color
|
Grey
|
|
Sample
|
Provide sample
|
|
Service
|
ODM/OEM
|
![]()
![]()
Specification
|
Sample Flow Rate
|
Range:50-150 ml/min; Accuracy:±20% ; Default: 75 ml/min
|
|
Principle of Operation
|
Non-dispersive infrared, dual wavelength, no moving parts
|
|
Initialization Time
|
Capnogram displayed in less than 4 seconds, at an ambient temperature of 25℃, full specifications within 2 mintues
|
|
CO2 measurement Range
|
0-114 mmHg ; 0-15% ; 0-15.2 kPa
|
|
CO2 Response Time
|
Includes Transport time and rise time.Less than 2 seconds, at 150 ml/min sampling flow
Less than 3 seconds, at 100 ml/min sampling flow
Less than 4 seconds, at 50 ml/min sampling flow
|
|
CO2 Resolution
|
0.2 mmHg @ 0-59 mmHg ; 0.5mmHg @60-114 mmHg
|
|
CO2 Accuracy
|
0-40 mmHg ± 2 mmHg41-70 mmHg ± 5% of reading
71-100 mmHg ± 8% of reading
101-114 mmHg ±10% of reading
Above 80 reaths per mintue ±12% of reading
* Note:Gas temperature at 25℃
|
|
CO2 Stability
|
Short Term Drift : Drift over three hours shall not exceed 1 mmHg maximumLong term Drift : Accuracy specification will be
maintained over a six hour period
|
|
CO2 Noise
|
RMS noise of the sensor shall be less than or equal to 0.5 mmHg at 5% CO2
|
|
Sampling Rate
|
100Hz
|
|
Respiration Rate Range
|
2 to 150 breaths per minute
|
|
Respiration Rate Accuracy
|
± 1 breath
|
|
Calibration
|
No Routine user calibration required
|
|
EtCO2 Calcuation
|
Method: peak of the expired CO2 waveform
|
|
Inspired CO2 Measurement
|
Method: Lowest reading of the CO2 waveform in the previous 20 seconds
|
|
Temperature
|
Operation : 0 to 40℃ ; Storage :-40 to 70℃ ; Transport :-40 to 70℃
|
|
Humidity
|
Operation :10 to 90 % RH, Non-condensing ; Storage :10 to 90 % RH, Non-condensing.
Transport: 10 to 90 % RH, Non-condensing.
|
|
Atmospheric Perssure
|
Operating :400-800 mmHg ; Storage :400-800 mmHg ; Transport :400-800 mmHg
|
|
Water Resisitance
|
IPX4-Splash-Proof- sensor only
|
|
Voltage Requirements
|
5 VDC ±5% ; Ripple not to exceed 100.0mV
|
|
Power Raing
|
< 1.2W Typical; < 2.0 W Warm-up
|
|
Sensor Weight
|
< 250 g
|
|
Sensor Size
|
Deep*Wide*High 102*62*38mm
|
|
Cable Length
|
75cm
|
Characteristics:
Long-term stability, non-dispersive infrared technology with two wavebands
Long service life, infrared black body light source with MEMS
Accurate Results, with built-in temperature, atmospheric pressure, compensation algorithm for the balance gas
Calibration-free, patented calibration algorithm without the need for calibration
Low sampling flow rate, the minimum sampling flow rate 50ml/min
Easy to operate, plug-in measure
Low cost in usage and storage
Specification
Respiration rate range: 2~150 breath/minute
Electric power supply: DC 5V
Power Consumption: 1.2W
Usage Environment: 0~40ºC, 10~90% RH, without condensation
Storage Environment: -40~70ºC, <90% RH, without condensation