ISO 5659 2 ASTM E662 NBS Smoke Density Test Chamber
Price:
Negotiable
MOQ:
1 set
Delivery Time:
25 days accrording the quantity
Brand:
YUYANG
Product Description
Product Overview
The ISO 5659-2 / ASTM E662 NBS Smoke Density Test Chamber measures the optical density of smoke generated when a test specimen of a specified thickness is exposed to a radiant heat source or open flame inside a sealed chamber. The ASTM E662 furnace delivers a heat flux of 2.5 W/cm² (2 kW/m²), while the ISO 5659 conical radiant heater allows adjustable irradiance levels of 10 kW/m², 25 kW/m², or 70 kW/m² to meet different testing requirements.
Applicable Standards
- ASTM E 662 — Standard Test Method for Specific Optical Density of Smoke Generated by Solid Materials.
- ISO 5659-2 — Plastics — Smoke generation — Part 2: Determination of optical density by a single-chamber test.
Structural Design and Construction
- The tester comprises a test cabinet, specimen holder assembly, radiant cone heater, ignition system, light source and measuring devices, plus the associated control components.
- Ergonomic structural and control design: the test box features a full-door design with a locking mechanism for one-time secure closure, streamlining operational procedures.
- The fully opening door allows easy loading and unloading of samples and convenient cleaning of the chamber interior.
- The front door is hinge-mounted with a viewing window, fitted with an opaque movable visor to prevent light from entering the sealed test box.
- The controller is mounted on the side of the cabinet for easy operation and remains visible even when the door is closed.
- Safety flameout board is simple to replace, ensuring safer operation.
- The test box interior is easy to clean and coated for chemical resistance.
Technical Specifications
| Parameter | Specification |
|---|---|
| Light transmittance measurement range | 0 ~ 100%; resolution 0.00001% |
| Measurement accuracy | Less than ±2.5% |
| Light source | 6.5 V incandescent lamp; 2200 ±100 K color temperature |
| Optical system | Linear transmittance response; measuring accuracy at least ±3% |
| Photomultiplier tube | Partial data acquisition precision 0.5% |
| Heater input temperature | 0 °C ~ 1000 °C; accuracy ±2 °C |
| Thermocouple | Imported K-type armoured thermocouple, accuracy 0.1 °C |
| Heat flux meter | Mouth thermopile type, range 0 ~ 80 kW/m²; accuracy 1%, repeatability 0.5% |
| Propane gas flow meter | Range 0 ~ 100 mL/min; minimum pressure 3.5 kPa ±1 kPa |
| Air flow meter | Range 0 ~ 1000 mL/min |
| Fan negative pressure in cabinet | At least 0.5 kPa |
| Enclosure pressure resistance | Greater than atmospheric pressure by at least 1.5 kPa |
| Water pressure gauge | 1.5 kPa (150 mm water column) |
| Wall temperature measurement range | 0 ~ 100 °C ±1 °C |
| Window heater power | 9 W |
| Pressure inside test chamber | < 1.5 kPa |
| Radiation power range | 0 ~ 50 kW/m² |
Specifications are subject to change for product improvement. Please confirm final configuration with our technical team.
Similar Products
Related Videos
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.
Company
DONGGUAN YUYANG INSTRUMENT CO., LTD
Location
B1505, FORTUNE PLAZA, No.1, Yuanmei Road, Nancheng District, Dongguan City, Guangdong Province, China
Contact Person
Jackson Sun