Lab Electrical Test Equipment Portable Multi Product Multimeter Calibration
GF6018A Portable Multi Product Multimeter Calibration Electronic Test Equipment with Voltage source & Current source
The equipment of clamp type multi product multimeter calibrator is compliance with national verification regulation: JJG124-2005 “Ammeter, voltmeter, power meter and resistance meter verification rules” and related national standards. 0.05 single-phase AC/DC standard source, can verify magnitude 0.2 and the following AC/DC voltmeter, ammeter, frequency meter, resistance meter. The source signals by using DSP and 16-bit high-speed analog-to-digital converters of controllable sine wave, the distorted wave signal source.The equipment comes with RS-232 interface, can be connected with PC to a semi-automatic verification system. It has precise interface and multi-functions including verification, storage and query.The equipment adopts color LCD screen, chart character display is clear, with high precision, stable and reliable, convenient operation and flexible characteristics.
Theory
Calibration of multimeter in measurement technology and metrology is the comparison of measurement values delivered by a device under test with those of a calibration standard of known accuracy. Such a standard could be another measurement device of known accuracy, a device generating the quantity to be measured such as a voltage, or a physical artefact, such as a metre ruler.
The outcome of the comparison can result in no significant error being noted on the device under test, a significant error being noted but no adjustment made, or an adjustment made to correct the error to an acceptable level. Strictly speaking, the term calibration means just the act of comparison, and does not include any subsequent adjustment.
The calibration standard of multimeter and clamp meter is normally traceable to a national standard held by a National Metrological Institute.
FEATURES
- It can be semi-automatic or manual inspection of analog multi meter, clamp type current meter, all kinds of electric meter, (voltmeter, ammeter, frequency meters, resistance meter) basic error
- The equipment can be generated with 2-31 harmonic wave, harmonic number, order, amplitude and harmonic phase of fundamental wave can be programmed
- Power amplifier working frequency range is 40Hz-1KHZ, has a good linear. Current amplifier is constant current source; voltage amplifier is constant voltage source. Because of light weight, this device is more suitable for field use
- With RS-232 interface, this device is controlled by PC through the PC software (optional), can be automatic or manual inspection, and the results are processing and management
- Has a non-volatile memory, it can store 500 pieces of the raw data of tested meter, for access and upload
PARAMETERS
| Electrical parameters | ||
|---|---|---|
| Power supply | Single phase AC 220V±10% or 110V±10%, 50/60HzAC | |
| AC Voltage output | ||
| range | 200mV, 400mV, 2000mV, 4000mV, 20V, 100V, 200V, 500V, 1000V | |
| Range 20V-1000V | Accuracy | ±(0.03%RD+0.02%FS) |
| Stability | ≤0.01% FS /60s | |
| Range 200mV-4000mV | Accuracy | ±(0.3%RD+0.2%FS) |
| Stability | ≤0.04% FS /60s | |
| Load capacity | 20VA | |
| Waveform distortion | ≤0.3% | |
| Adjustable Range | 0-120% FS (1000V not included) | |
| Fineness | 5×10-5 | |
| AC Current Output | ||
| Range | 2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A | |
| Range 0.2A-20A | Accuracy | ±(0.03%RD+0.02%FS) |
| Stability | ≤0.01% FS /60s | |
| Range 2mA-40mA | Accuracy | ±(0.3%RD+0.2%FS) |
| Stability | ≤0.04% FS /60s | |
| Load capacity | 20VA | |
| Waveform distortion | ≤0.3% | |
| Adjustable Range | 0-120% FS | |
| Fineness | 5×10-5 | |
| Frequency Output | ||
| Range | 45-65Hz | |
| Fineness | 0.001Hz | |
| Accuracy | 0.01Hz | |
| Harmonic Wave | ||
| Times | 2-31 | |
| Extent | 0-20% | |
| Each harmonic phase fineness | 0.01·N (N is harmonic order) | |
| DC voltage output | ||
| Range | 200mV, 400mV, 2000mV, 4000mV, 20V, 40V, 100V, 200V, 400V, 1000V | |
| Load capacity | 20W | |
| Adjustable range | 0-110% FS (1000V not included) | |
| Adjustable fineness | 5×10-5 | |
| Stability | ≤0.01% FS /60s (peak-peak value) | |
| Accuracy | ±(0.03%RD+0.02%FS) | |
| Ripple wave | ≤1% FS | |
| DC current output | ||
| Range | 20uA, 40uA, 200uA, 400uA, 2mA, 4mA, 20mA, 40mA, 0.2A, 1A, 2.5A, 5A, 10A, 20A | |
| Load capacity | 20W | |
| Adjustable Range | 0-110% FS (1000V not included) | |
| Adjustable Fineness | 5×10-5 | |
| Stability | ≤0.01% FS /60s (peak-peak value) | |
| Accuracy | ±(0.06%RD+0.04%FS) | |
| Ripple wave | ≤1% FS | |
| Resistance box | ||
| Range | 0-200Ω, 200-400Ω, 0-2kΩ, 2kΩ-4kΩ, 0-20kΩ, 20kΩ-40kΩ, 0-200kΩ, 200kΩ-400kΩ, 0-2MΩ, 2MΩ-4MΩ | |
| Fineness | 0.02% | |
| Range 0-2kΩ, 0-20kΩ, 0-200kΩ | ||
| Range 0-2kΩ | Adjustable Range | 0-100%FS |
| Accuracy | ±(0.12%RD+0.08%FS) | |
| Range 0-200Ω,0-2MΩ | Adjustable Range | 0-100% FS |
| Accuracy | ±(0.3%RD+0.2%FS) | |
| Range 2kΩ-4kΩ, 20kΩ-40kΩ, 200kΩ-400kΩ | ||
| Range 2kΩ-4kΩ | Adjustable Range | 5%-100%FS |
| Accuracy | ±(0.12%RD+0.08%FS) | |
| Range 200-400Ω, 2MΩ-4MΩ | Adjustable Range | 5%-100% FS |
| Accuracy | ±(0.3%RD+0.2%FS) | |
| Clamp meter range | ||
| 200A | Accuracy | ±0.2% |
| 600A | Accuracy | ±0.5% |
| 1000A | Accuracy | ±1% |
| Dot frequency square wave output | ||
| Range | 1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz | |
| Square wave amplitude | 4.5V | |
| Setting accuracy | 5×10-5 | |
| Mechanical parameters | ||
| Dimension (L×W×H) (mm) | 440x360x160 | |
| Weight (kg) | 12 | |
| Environmental conditions | ||
| Operating temperature | 20°C±10°C | |
| Relative humidity | ≤85%RH | |