Lab Reference Meter Test Equipment Three Phase With 128st Harmonics Measurement
GF333B three phase portable reference meter test equipment with 128st harmonics measurement
GF333B three phase portable reference meter is a reference standard with the characteristics of wide-range, multifunction and high-precision. It has accuracy class of 0.02. By adopting techniques of DSP, embedded system and automatic temperature balancing and other compensation, it has not only advantages of light weight, small size, high precision, high performance, powerful function, original interface, ease-to-use, but also work stability.The product can be widely used in fields of electric energy measuring, electric energy laboratories and other relevant industry, not only in laboratories but also at the industrial field.
FEATURES
- High accuracy up to 0.05%
- Wide testing range: voltage 5V-480V, current 1mA-120A
- Testing type: 3 phase 4 wire, 3 phase 3 wire
- Testing mode: active power, reactive power, apparent power
- Harmonic analysis function
- Waveform display function
- Vector diagram function
- Energy accumulating function
- Internal pulse input port, testing meter’s error directly
PARAMETERS
| Electrical parameters | |
|---|---|
| Accuracy | 0.05% |
| Voltage supply | 220V±10% or 110V±10%, 50/60Hz |
| Voltage measurement | |
| Range | 5V-480V |
| Error | ±0.02% (50V-480V), ±0.05% (5V-50V) |
| Display range | 5.000000V-480.0000V |
| Harmonic | 2nd-63st |
| Current measurement | |
| Range | 1mA-120A |
| Error | ±0.02% (0.2A-120A), ±0.05% (1mA-0.2A) |
| Display range | 1.000000mA-120.0000A |
| Power measurement | |
| Active power | ±0.02% (0.2A-120A) ±0.05% (0.01A-0.2A) ±0.1% (0.001A-0.01A) |
| Reactive power | ±0.05% (0.2A-120A), ±0.1% (0.001A-0.2A) |
| Apparent power | ±0.05% (0.2A-120A), ±0.1% (0.001A-0.2A) |
| Energy error | |
| Active energy | ±0.02% (0.2A-120A) ±0.05% (0.01A-0.2A) ±0.1% (0.001A-0.01A) |
| Reactive energy | ±0.05% (0.2A-120A), ±0.1% (0.001A-0.2A) |
| Apparent energy | ±0.05% (0.2A-120A), ±0.1% (0.001A-0.2A) |
| Phase measurement | |
| Range | 0°-360° |
| Error | ±0.005° |
| Display range | 0.0001°-359.99° |
| Frequency measurement | |
| Range | 45-65Hz |
| Display range | 45.0000-65.0000 |
| Accuracy | 0.002Hz |
| Power Factor measurement | |
| Range | -1.0 ~ 0 ~ +1.0 |
| Resolution | 0.0001 |
| Accuracy | 0.0005 |
| Energy pulse | |
| High frequency output(CH) | 12000Hz |
| Low frequency output(CL) | 5000Hz |
| Pulse ratio | 1:1 |
| Output level | 5V |
| Input level | 5V |
| Input frequency | Max. 1MHz |
| Mechanical parameters | |
| Dimensions (W×H×D) (mm) | 480×138×486 |
| Weight (kg) | 10 |
| Environmental conditions | |
| Ambient temperature | -10°C to 45°C |
| Influence of external fields | ≤0.05 %/mT |
| Relative humidity | 10%-80% |
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.