High Voltage Insulating Boots & Gloves Test Equipment
This specialized test equipment is designed for evaluating the leakage current, insulation aging, and power frequency performance of insulating gloves and boots. The system can simultaneously test 3 pairs of protective gear, providing individual leakage current readings for each unit to quickly identify non-compliant items.
Key Features
High-speed testing capability for improved operational efficiency
Direct display of voltage and leakage current with automatic identification of unqualified products
Simultaneous testing of 3 pairs of gloves or boots with individual current readings
Integrated trundles for convenient mobility between workstations
Technical Specifications
Input Voltage AC 220V, 50Hz
Output Voltage 0-30kV or 50kV
Capacity 3kVA or 5kVA
Measurement Quantity 6 units per test cycle
Measurement Data Display and print voltage and leakage current for each unit
Display System LCD display with integrated print function
Accuracy Voltage: ±2% (±3 bits), Current: ±2% (±3 bits)
Total Weight Approximately 80-100kg
High voltage insulating boots and gloves test equipment front view showing control panel and testing chambers Close-up view of test equipment interface and electrode configuration for safety gear testing
Operational Guidelines
Proper operation of the insulation boot and glove withstand voltage test device requires adherence to specific safety protocols and maintenance procedures.
Pre-Use Requirements
Test transformer insulation resistance before operation (input > 2MΩ, output > 10MΩ)
Verify electrical component contacts and protection system functionality
Clean insulation supports, electrode rods, and water tank components with alcohol
Testing Procedures
Automatic power cutoff when leakage current exceeds specifications or material breaches occur
Audible alarm system for non-compliant insulation gear
Voltage adjustment capability via control panel during power fluctuations
Post-Test Maintenance
Thorough drying of all components using cotton materials
Storage in dry, well-ventilated environments for extended periods
Important: Operation must be conducted by qualified professionals following established safety protocols. Ensure work areas are free from explosive gases, corrosive media, and other insulation-affecting contaminants.
Request Custom Quote

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Wuhan GDZX Power Equipment Co., Ltd
Location No.1 Fenghuangyuan Second Road, Jiangxia District, Wuhan City, Hubei Province, China
Contact Person Annie

Request A Quote

Please check your email address.
Your message must be at least 20 characters.