KTG113-W Trace Oxygen Analyzer for On-line Analysis of Magnetic Materials 400 ± 10 ml/min Sample Gas Flow Rate
The KTG113-W trace oxygen analyzer is used for the on-line analysis of magnetic materials
Introduction
The KTG113-W trace oxygen analyzer is a new type of intelligent industrial on-line analyzer developed by combining the electrochemical detection principle, imported high-performance electrochemical sensors, and new microcomputer technology.
Application Areas
On-line analysis of oxygen content in nitrogen production from air separation, chemical processes, protective gases in high-temperature sintering furnaces for magnetic materials, protective gases in the electronics industry, as well as in the metallurgy, glass, and building materials industries.
Instrument Features
- It uses imported fuel cell sensors, which feature long service life, high accuracy, and fast response.
- With electronic flow control, there is no blockage problem.
- It can be used to measure trace oxygen in reducing gases such as high-purity hydrogen.
- It comes with a precision filter, which greatly extends the service life of the sensor.
- It has a fully Chinese man-machine dialogue menu, making operations intuitive and convenient.
- It has a large 320×240 dot-matrix LCD screen, which can simultaneously display parameters such as oxygen content, flow rate, date, time, and the internal temperature of the instrument.
- It has automatic temperature compensation to eliminate the influence of ambient temperature.
- It has a timed automatic storage function, allowing users to view stored data and curves at any time.
- The upper and lower limit control points can be set arbitrarily within the full measurement range.
- It is equipped with a standard RS232 or RS485 communication port, enabling two-way communication with a computer.
Technical Parameters
Measurement Range: (0.00 - 10/100/1000/2000) × 10⁻⁶ (The range is variable. Please specify when placing an order).
Maximum Allowable Error: ±2% FS (0.00 - 10) × 10⁻⁶; ±1% FS (10 - 1000/2000) × 10⁻⁶.
Stability: Zero drift: ±1% FS/7d; Span drift: ±1% FS/7d.
Repeatability: ±1% FS.
Sample Gas Flow Rate: (400 ± 10) ml/min.
Response Time: T90 ≤ 15 seconds.
Resolution: 0.01 × 10⁻⁶.
Sample Gas Pressure: 0.05 MPa ≤ Inlet pressure ≤ 0.1 MPa.
Analog Output: (4 - 20) mA standard signal.
Digital Output: Standard RS485 communication port, enabling two-way communication with a computer.
Contact Output: Dual relay outputs: 220VAC, 1A or 24VDC, 2A.
Operating Environment: Temperature: -15°C - +45°C; Humidity: ≤90% RH.
Operating Power Supply: (220 ± 22) VAC, (50 ± 5) Hz.
Overall Dimensions: 480mm (Width) × 136mm (Height) × 370mm (Depth).
Installation Dimensions: 442mm (Width) × 134mm (Height).
Weight: Approximately 7.6 kg.
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.