KTG141 Explosion-proof Infrared Gas Analyzer Advanced Technology For Inflammable And Explosive Gas Analysis
The KTG141 Explosion - proof Infrared Gas Analyzer is used for the analysis of inflammable and explosive gases.
Introduction
The KTG141 Explosion - proof Infrared Gas Analyzer adopts imported non - dispersive infrared sensors and advanced digital processing technology to achieve continuous, automatic, rapid and online detection of various gases such as CO, CO₂, CH₄, NH₃ and C₆H₁₄. Compared with similar products, the most prominent features of this product are high stability, high reliability and high selectivity.
Application Areas
The KTG141 has a flame - proof explosion - proof structure and can be widely and safely applied in oil refining, chemical industry, natural gas and other occasions where there are inflammable and explosive gases.
Instrument Features
- It adopts imported infrared sensors, which have a long service life, high sensitivity and fast response speed.
- It is equipped with a precision filter, which greatly prolongs the service life of the sensor.
- It has a full - Chinese man - machine dialogue menu, which is intuitive and convenient to operate.
- The instrument adopts a high - performance embedded integrated touch - screen TFT true - color display, which can simultaneously display the measured value, historical curve, range, status and various parameter setting interfaces. Various instrument parameters can be configured, calibrated and tested through the full - Chinese touch - screen menu or remote control. The historical records and curves can be set according to time.
- It has control contact output, over - limit self - alarm (buzzer), and the control mode can be set at will. The standard current output is (4 - 20) mA.
Technical Parameters
Measuring range:
| Measured Gas | Minimum Range | Maximum Range% |
| CO | (0 - 100)×10⁻⁶ | (0 - 100)% |
| CO₂ | (0 - 20)×10⁻⁶ | (0 - 100)% |
| NH₃ | (0 - 500)×10⁻⁶ | (0 - 100)% |
| CH₄ | (0 - 100)×10⁻⁶ | (0 - 100)% |
Maximum allowable error: ±1.5% FS.
Stability: Zero - point drift: ±1% FS/7d; Span drift: ±1% FS/7d.
Sample gas flow rate: (400 ± 20) mL/min.
Response time: τ90 ≤ 20S (when the gas passes directly through the gas chamber).
Sample gas pressure: 0.05MPa ≤ inlet pressure ≤ 0.25MPa.
Output signal: Optional (4 - 20) mA or (0 - 10) mA.
Contact capacity: 220V AC, 1A; 24V DC, 2A.
Working power supply: (220 ± 22) V AC, (50 ± 5) Hz.
Working environment: Temperature: - 10℃ - + 45℃; Humidity: ≤90% RH.
Explosion - proof grade: ExdII CT6.