KGD WQ 508 Partial Flow Particulate Matter Sampling System With Dilution Exhaust Gas Flow Control Range 35 100 SLM

Price Negotiable
Price: Negotiable
MOQ: 1PCS
Delivery Time: 5-8 work days
Brand: KACISE
Product Description
KGD WQ 508 Partial Flow Particulate Matter Sampling System With Dilution Exhaust Gas Flow Control Range 35-100 SLM
Product Attributes
Attribute Value
Applicable Engine Range Engines of any size, at any speed, and with exhaust pipes of various specifications
Sampling Flow Measurement Accuracy ≤ ±4% (Dilution ratio ≤ 15:1)
Dilution Air Flow Control Range 35 - 100 SLM
Dilution Exhaust Gas Flow Control Range 35 - 100 SLM
Dilution Air Flow Control Accuracy ≤ ±1%
Dilution Exhaust Gas Flow Control Accuracy ≤ ±1%
Product Description

KGD-WQ-508 Partial-flow Particulate Matter Sampling System

Product Introduction

The KGD-WQ-508 Partial-flow Particulate Matter Sampling System is an automated device that mixes clean air with a small amount of raw exhaust gas sampling airflow to form diluted exhaust gas, and collects particulate matter in the diluted exhaust gas through a filter paper. This device uses a high-frequency flow control strategy to quickly and accurately control the sampling flow rate, making it change proportionally and synchronously with the total flow rate of the engine's raw exhaust gas. Thus, it can quantitatively analyze the PM concentration of exhaust gas particulate matter under the transient operating conditions of the engine.

KGD-WQ-508 Partial-flow Particulate Matter Sampling System
Product Features
  • Has two operating modes: proportional dilution and fixed-ratio dilution
  • Has an automatic filter paper switching function
  • Has a one-key automatic operation function for flow calibration and linearity check
  • Has a constant-temperature control system for dilution air and diluted exhaust gas
Technical Parameters
Parameter Details
Applicable Engine Range Engines of any size, at any speed, and with exhaust pipes of various specifications
Sampling Flow Measurement Accuracy ≤ ±4% (Dilution ratio ≤ 15:1)
Dilution Air Flow Control Range 35 - 100 SLM
Dilution Exhaust Gas Flow Control Range 35 - 100 SLM
Dilution Air Flow Control Accuracy ≤ ±1%
Dilution Exhaust Gas Flow Control Accuracy ≤ ±1%
Filter Paper Size 47mm diameter or 70mm
Pre-Heated Temperature of Filter Paper 47 - 55°C
Gas Flow Rate through Filter Paper 35 - 100 cm/s

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Xi'an Kacise Optronics Co.,Ltd.
Location 2nd Building, Tianyuan International Mansion, High-tech Zone, Xi'an City, Shaanxi Province, China
Contact Person Evelyn Wang

Request A Quote

Please check your email address.
Your message must be at least 20 characters.