Dwyer Series 2000 Magnehelic Differential Pressure Gage (Model 2300-60PA)

Price Negotiable
Price: Negotiable
MOQ: 1PCS
Delivery Time: 5-8 work days
Brand: Dwyer
Product Description
Dwyer Series 2000 Magnehelic Differential Pressure Gage (Model 2300-60PA)
Attribute Value
Weight 0.5kg
Warranty 12 months
Brand Name Dwyer
Accuracy 0.0005
Range 2000-60pa
The Dwyer Series 2000 Magnehelic Differential Pressure Gage (Model 2300-60PA) is a versatile, low-cost solution for measuring positive, negative, or differential pressures in air and non-corrosive gas systems.
Dwyer Series 2000 Magnehelic Differential Pressure Gage
Key Features
Precision Measurement:
  • Range: 30-0-30 Pa with 1.0 Pa minor divisions, calibrated for vertical scale position
  • Accuracy: ±2% of full scale (±1% with -HA high-accuracy option) at 70°F (21.1°C)
  • Measures positive, negative (vacuum), or differential pressures in air and non-corrosive gases
Durable & Rugged Design:
  • Die-cast aluminum case with iridite dip and baked gray finish, passing 168-hour salt spray tests
  • IP67 enclosure rating for weatherproofing and protection against dust/water immersion
  • Overpressure relief plug opens at ~25 psig (1.72 bar) to prevent damage (standard models)
Functional Versatility:
  • No power required, suitable for isolation/hospital environments and temporary units
  • 1/8" female NPT pressure taps (side and back) for flexible installation
  • Applications include filter monitoring, air velocity with Pitot tubes, blower vacuum, and room pressure indication
User-Friendly Design:
  • 4" diameter dial with clear acrylic cover for undistorted, far-viewing readability
  • Red-tipped aluminum pointer on a precision litho-printed scale
  • Zero adjustment screw accessible without removing the cover; O-ring sealed for pressure tightness
Technical Specifications:
  • Pressure limits: -20 in Hg to 15 psig (-0.677 to 1.034 bar); optional MP/HP up to 35/80 psig
  • Temperature range: 20-140°F (-6.67-60°C); -20°F (-28°C) with low-temperature option
  • Weight: 1 lb 2 oz (510 g); includes mounting adapters and plugs
Safety & Compliance:
  • Meets EU RoHS II Directive (2011/65/EU); explosion-proof options via Series AT-2000 for ATEX zones
  • Compatible with hydrogen when ordered with Buna-N diaphragm (pressures

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Xi'an Kacise Optronics Co.,Ltd.
Location 2nd Building, Tianyuan International Mansion, High-tech Zone, Xi'an City, Shaanxi Province, China
Contact Person Evelyn Wang

Request A Quote

Please check your email address.
Your message must be at least 20 characters.