Bacteria Cultivation Anaerobic Incubator With Deoxidization Catalyst 50~90%RH
Stainless Steel LAI-D2Anaerobic Workstation Glove Boxes With Deoxidization Catalyst
Not all Chambers Are Anaerobic
You may see other chambers around the lab, but they aren’t all for anaerobic use. Don’t confuse the following with anaerobic chambers:
- Dry boxes
- Full humidity control
- Oxygen tissue culture glove boxes
- Oxygen animal study glove boxes
Anaerobic chambers are pretty intimidating if you have never seen or worked in one. They can be tricky and there is a huge learning curve when working with gasses, regulators, seals, vacuums, redox states, and proteins precipitating or even sometimes evaporating. However, if you take good care of the anaerobic chambers they can last through several PhDs and lead to some interesting science.
Specifications:
| Model | LAI-D2 |
| Time for creating anaerobic state in sample chamber | <5 minutes |
| Time for creating anaerobic state in operation chamber | <1hr |
| Anaerobic environment maintenance time | >12hrs(when no supply of mixed gas) |
|
Temperature Range
|
RT+3~60℃ |
| Temperature Stability | <±0.3℃ |
| Temperature Uniformity | <±1℃ |
| CO2 Range | 0~20% |
| CO2 Control Accuracy | ±0.1%(imported sensor) |
| Humidity Control Range | 50~90%RH |
| Humidity Fluctuation | ±3%RH |
| Power Rating | 1500W |
| Power Supply | AC 220V,50HZ(can customized) |
| Interior Chamber Size (W×D×H)cm | 42×29×47.5 |
| Operation Chamber Size (W×D×H)cm | 95×67×75 |
| Sampling chamber size (W×D×H)cm | 40×30×32 |
| Shell material | All 304 stainless steel |