70W Zetron OPC-L2 Oil Particle Counter Photo Resistive Method

Price Negotiable
Price: Negotiable
MOQ: 1 pcs
Delivery Time: 14 work days
Brand: Zetron
Product Description

OPC-L2 Oil Particle Counter

 

OVERVIEW
 

OPC-L2 Portable Oil Particle Counter adopts Laser Extinction Technology, is designed to measure the particles’ size and quantity in oil liquid. It is Suitable for tests of pollution degree of solid particles in hydraulic oil, lubricating oil, fire resistant oil, insulating oil, turbine oil, organic liquid and polymer solution, etc. So our instrument can be used in aviation, spaceflight, electric power, petroleum, chemical industry, transportation, port, metallurgy, machinery, automobile manufacturing and many other fields.4.1 Power Choice: External power or Battery.

 
FEATURES
 

* Adopting the counting principle of photoresist (shading) method formulated by the International Hydraulic Standard Committee
* High-precision laser sensor with wide testing range, stable performance, low noise and high resolution.
* Adopt high pressure syringe pump sampling mode, can set the sampling volume, stable sampling speed, high sampling accuracy.
* Positive and negative pressure combination of sampling system, can realize the sample degassing, suitable for different viscosity of the inspection products testing.
* Built-in pressure sensor, you can set the pressure value, and automatically determine the air pressure in the chamber to ensure safety.
* Built-in air purification system to ensure that the test is not contaminated
* Built-in multiple calibration curves, compatible with all common domestic and international standards for calibration.
* Built-in GJB-420A, GJB-420B, NAS1638, GB/T14039, ISO4406, SAE4059cpc, SAE4059F, SAE749D, ГOCT17216, QC/T29104, JB/T9737, DLT432, HH005-2018 and other more than ten commonly used standards. One test can give the data results under all built-in standards.
* Supports customized standard testing and grade determination
 

APPLICATION SCENES

 
OPC-L2 oil particle counter can be widely used in aviation, aerospace, electric power, petroleum, chemical industry, transportation, port, metallurgy, machinery, automobile manufacturing and other fields.
 

PRODUCT PARAMETERS

 

Light source
Semiconductor laser
Particle size range
0.8μm~600μm (calibrated according to 1~100um or 4~70um(c))
Detection channels
8~64 channels optional, arbitrary setting of particle size size
Sensitivity
0.8μm (ISO4402) or 3μm(c) (GB/T18854, ISO11171)
Resolution
<10% (GB/T18854, ISO11171)
Repeatability
RSD<2%
Sample detection viscosity
≤650cSt (viscosity is too large can be combined with heating or dilution method detection)
Sampling volume
0.2~1000mL, interval 0.1mL
Sampling accuracy
better than ± 0.5%
Sampling speed
5 ~80mL/min
Pneumatic chamber
Positive and negative air pressure chamber device, to realize sample degassing and high viscosity sample detection
Maximum vacuum of air pressure chamber
-0.08Mpa
Maximum positive pressure of air pressure chamber
0.8 Mpa
Maximum particle concentration
12000~40000 particles/mL
Power supply
110~245V AC, 50Hz, 70W
Overall Dimension
340mm*410mm*650mm
Net weight
25kg

 

 
 
 
 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Beijing Zetron Technology Co., Ltd
Location Room 20E, Block 1-A, Oriental Ginza, No. 48 Dongzhimenwai Avenue, Dongcheng,100007, Beijing, China
Contact Person Lisa Niu

Request A Quote

Please check your email address.
Your message must be at least 20 characters.