Anhui Wanyi Science and Technology Co., Ltd.
                                                                                                           
Verified Supplier
23 Years
Since 2003
Menu

DAD Intelligent Liquid Chromatography Instrument Hplc Testing Equipment 1800L/mm

Price Negotiable
Price: Negotiation
MOQ: 1
Delivery Time: 35 days
Brand: Wayeal
Product Description

Intelligent High Performance Full-controlled hplc Liquid Chromatography Instrument

Introduction of HPLC Liquid Chromatography Instrument

High performance liquid chromatograph (HPLC) is consists of reservoir, pump, sample injector, chromatographic column, detector and recorder. And it is mainly used to analyze less-volatile and thermal-unstable organic compounds with high boiling point and large molecular weight.

Parameters of HPLC Liquid Chromatography Instrument

Pump

Flow rate range 0.001-10.000mL/min( increments :0.0001 mL/min)
Max operating Pressure 45MPa
Pressure pulsation <0.1MPa( Test condition: 10MPa , at 1.0ml/min methanol)
Flow accuracy ±0.2%( Test condition: 10MPa , at 1.0ml/min methanol)
Flow precision RSD≤0.06%( Test condition: 10MPa , at 1.0ml/min methanol)
Gradient accuracy ±1%
Seal wash With programmable on-line seal wash module

Autosampler

Injection Volume Precision <0.25%RSD( full loop) ; <0.5%RSD( partial loop) ; <1%RSD( no waste)
Linearity >0.999
Cross Contamination(Carry Over) <0.01%
Sample Capacity 120 samples( 1.5ml vials)
Syringe 250 μL( Standard, with 500μL,1000μL, 2500μL optionals)
Sample Loop 20μL( Standard, with 50μL, 100μL, 200μL optional)
Optional sample chamber cooler 4 °C ~ 30 °C(4 °C lowest)
Injection volume range 1-100 μL
Injection mode Full loop injection,partial loop injection,no waste injection

DAD Detector

Noise 1.5×10-5Au
Drift ≤1×10-3Au/h
LOD 5.0×10-9g/mL (naphthalene/methanol)
Wavelength range 190-800n
Wavelength accuracy ≤±1 nm
Wavelength reproducibility ≤±0.1 nm
Linearity ≥1.0×104
Extend functions Wavelength bunching etc.

FLD Detector

Noise 1×10-5 Fu
Drift ≤2×10-4Fu/30min
LOD 5.0×10-10g/mL (naphthalene/methanol)
Wavelength range 200-650nm
Wavelength accuracy ±2 nm
Wavelength reproducibility ±0.2 nm
Linearity ≥1.0×104

Performance of Liquid Chromatography

1. Mainstream split design, complete system testing function, to meet most testing needs;
2. Customized acetonitrile-resistant check valve;
3. The new cam curve design adopts electronic pulsation suppression technology and solvent
compression compensation to further reduce pressure fluctuations and ensure higher repeatability of test results;
4. Standard configuration of plunger rod online cleaning function;
5. 1800L/mm grating imported from France;
6. Optional UV, ELSD, FLD, DAD detectors from our company;
7. Fully intelligent counter-control chromatography workstation

 

 

 

 

 

 

 

 

 

 

 

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Anhui Wanyi Science and Technology Co., Ltd.
Location No. 8, Wenqu Road, High-tech Zone, Hefei, Anhui
Contact Person Annie

Request A Quote

Please check your email address.
Your message must be at least 20 characters.