4O2-LQ Gas Oxygen Sensor Instruments And Meters For Oxygen Concentration Detection

Price Negotiable
Price: Negotiable
MOQ: 1
Delivery Time: 3-5work days
Product Description

 

                                                          4O2-LQ Gas oxygen sensor

 

Basic Information:

  • Measurement Range: Typically 0–30% vol. O₂ (also available in specifications like 0–25% vol. O₂).
  • Model Examples: CITY brand 4-O₂ (Lq), product codes such as 2112B0231400 (CLE-0231-400).
  • Application Fields: Instruments and meters for oxygen concentration detection.

Technical Features:

  • Technology Principle: Adopts solid-state electrochemical technology with a three-electrode advanced design and standard liquid electrolyte, ensuring high stability.
  • Dimensions: Standard 4-series size.

Performance Parameters:

  • Output Type: Analog output.
  • Accuracy: Specific accuracy indicators may vary by manufacturer and batch, but generally meet the precision requirements for oxygen concentration detection in relevant scenarios.
  • Response Time: Quickly responds to changes in oxygen concentration (specific time parameters refer to the product manual).
  • Stability: Features good long-term stability due to its advanced design, maintaining accurate measurement performance over time.

Other Information:

  • Compatibility: Some 4O2-LQ sensors are compatible with related devices from other brands, such as certain models from RAE.
  • Working Environment:
  • Temperature: Typically operates within a wide range, e.g., -35°C to +70°C
  • Humidity: Usually 5%–100% RH (specific requirements depend on product specifications).

 

Sensor Type Electrochemical oxygen sensor
Measurement Range 0–30% vol. O₂ (standard); optional 0–25% vol. O₂
Product Code e.g., 2112B0231400 (CLE-0231-400)
Sensing Principle Three-electrode solid-state electrochemical with liquid electrolyte
Output Signal Analog current/voltage (e.g., 0.1–1.0 mA in air)
Accuracy ±1% of full scale (25°C, 1 atm)
Response Time (T90) ≤15 seconds (to 90% of final value)
Zero Drift <200 μA/year (in clean air)
Linearity >99% within 0–30% O₂ range
Operating Temperature -35°C to +70°C (non-condensing)
Storage Temperature -20°C to +50°C (recommended: 5°C–25°C)
Humidity Range 5%–100% RH (condensing allowed for short periods)
Pressure Range 80–120 kPa (0.8–1.2 atm)
Dimensions Φ20.0 mm × 16.8 mm (diameter × height)
Weight ~16 g
Load Resistance 10–100 Ω (recommended: 47 Ω)
Service Life ≥24 months (in clean air, 25°C)
Storage Life 6 months (original packaging)
Cross-Sensitivity - CO (100 ppm): <0.1% O₂ error
- H₂S (10 ppm): <0.5% O₂ error
Certifications RoHS compliant, CE marked

 

 

 

 

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company ShenzhenYijiajie Electronic Co., Ltd.
Location Room 3306AB 33rd Floor, SEG Plaza,Huaqiang North Road, Futian District, Shenzhen Guangdong Province 518028
Contact Person Xu

Request A Quote

Please check your email address.
Your message must be at least 20 characters.