Keithley 2460 100V 7A 100W Source Measure Unit SMU Instrument with Touchscreen
The 2460 High Current SourceMeter® Source Measure Unit (SMU) Instrument brings advanced Touch, Test, Invent® technology right to your fingertips. It combines an innovative graphical user interface (GUI) with capacitive touchscreen technology to make testing intuitive and minimize the learning curve to help engineers and scientists learn faster, work smarter, and invent easier.
With its 7A DC and pulse current capability, the 2460 is optimized for characterizing and testing high power materials, devices, and modules such as silicon carbide (SiC), gallium nitride (GaN), DC-DC converters, power MOSFETs, solar cells and panels, LEDs and lighting systems, electrochemical cells and batteries, and much more. These new capabilities, combined with Keithley's decades of expertise in developing high precision, high accuracy SMU instruments, will make the 2460 a "go-to instrument" for high current applications in the lab and in the rack for years to come.
- One tightly coupled instrument that combines capabilities from analyzers, curve tracers, and I-V systems at a fraction of their cost
- Wide coverage up to 105 V, 7 A DC/7 A pulse, 100 W max.
- Five-inch, high resolution capacitive touchscreen GUI
- 0.012% basic measure accuracy with 6½-digit resolution
- Source and sink (4-quadrant) operation
- Four "Quickset" modes for fast setup and measurements
- Context-sensitive help function
- Front panel input banana jacks; rear panel input mass termination screw connections
- 2460 SCPI and TSP® scripting programming modes
- Front-panel USB 2.0 memory I/O port for transferring data, test scripts, or test configurations
The 2460 features a five-inch, full-color, high resolution touchscreen that supports intuitive operation, helps operators become familiar with the instrument quickly, and optimizes overall speed and productivity. A simple icon-based menu structure reduces the number of steps required to configure a test by as much as 50 percent and eliminates the cumbersome multi-layer menu structures typically used on soft-key instruments.
Built-in, context-sensitive help supports intuitive operation and minimizes the need to review a separate manual. These capabilities, combined with the 2460's high versatility, simplify its operation in both basic and advanced measurement applications, regardless of the user's previous experience in working with SMU instruments.
- Max. Output Power: 100 W, four-quadrant source or sink operation
- Source Limits: Vsource: ±7.35 V (≤7 A range), ±10.5 V (≤5 A range), ±21 V (≤4 A range), ±105 V (≤1 A range)
- Source Limits: Isource: ±7.35 A (≤7 V range), ±5.25 A (≤10 V range), ±4.2 A (≤20 V range), ±1.05 mA (≤100 V range)
- Overrange: 105% of range, source and measure
- Regulation - Voltage: Line: 0.01% of range. Load: 0.01% of range + 100 µV
- Regulation - Current: Line: 0.01% of range. Load: 0.01% of range + 100 pA
- Source Limits: Voltage Source Current Limit: Bipolar current limit set with single value. Min. 10% of range
- Source Limits: Current Source Voltage Limit: Bipolar voltage limit set with single value. Min. 10% of range
- V-Limit / I-Limit Accuracy: Add 0.3% of setting and ±0.02% of reading to base specification
- Overshoot - Voltage Source: <0.1% typical (full scale step, resistive load, 20 V range, 10 mA I-Limit)
- Overshoot - Current Source: <0.1% typical (1 mA step, RLoad = 10 kΩ, 20 V range)
- Range Change Overshoot: Overshoot into a fully resistive 100 kΩ load, 10 Hz to 20 MHz BW, adjacent ranges: <250 mV typical
- Output Settling Time: Time required to reach 0.1% of final value after command is processed and output slew, 20 V range, 100 mA I-Limit: <200 µs typical
- Maximum Slew Rate: 1 V per µs, 100 V range, 100 mA limit into a 20 kΩ load (typical). 0.6 V per µs, 20 V range, 100 mA limit into a 20 kΩ load (typical)
- Over Voltage Protection: User selectable values, 5% ±0.5 V tolerance. Factory default = none
- Voltage Source Noise: 10 Hz-20 MHz (RMS): <4.5 mV typical into a resistive load
| Range | Max. Current | Source Resolution | Source Accuracy (23° ± 5°C) 1 Year ±(% setting + volts) | Source Noise (RMS) (<10 Hz) | Measure Resolution | Input Resistance | Measure Accuracy (23° ± 5°C) 1 Year ±(% rdg. + volts) |
|---|---|---|---|---|---|---|---|
| 200.0000 mV | 7.35 A | 5 μV | 0.015% + 200 µV | 1 µV | 100 nV | >10 GΩ | 0.012% + 200 µV |
| 2.000000 V | 7.35 A | 50 µV | 0.015% + 300 µV | 10 µV | 1 µV | >10 GΩ | 0.012% + 300 µV |
| 7.000000 V | 7.35 A | 250 µV | 0.015% + 2.4 mV | 100 µV | 1 µV | >10 GΩ | 0.015% + 1 mV |
| 10.00000 V | 5.25 A | 250 µV | 0.015% + 2.4 mV | 10.00000 V | 10 µV | >10 GΩ | 0.015% + 1 mV |
| 20.00000 V | 4.20 A | 500 µV | 0.015% + 2.4 mV | 100 µV | 10 µV | >10 GΩ | 0.015% + 1 mV |
| 100.0000 V | 1.05 A | 2.5 mV | 0.015% + 15 mV | 1 mV | 100 µV | >10 GΩ | 0.015% + 5 mV |
| Range | Max. Voltage | Source Resolution | Source Accuracy (23° ±5°C) 1 Year ±(% rdg. + amps) | Source Noise (RMS) (<10 Hz) | Measure Resolution | Voltage Burden 6 | Measure Accuracy (23° ±5°C) 1 Year ±(% rdg. + amps) |
|---|---|---|---|---|---|---|---|
| 1.000000 μA | 105 V | 50 pA | 0.025% + 1 nA | 40 pA | 1 pA | <100 µV | 0.025% + 700 pA |
| 10.00000 μA | 105 V | 500 pA | 0.025% + 1.5 nA | 40 pA | 10 pA | <100 µV | 0.025% + 1 nA |
| 100.0000 μA | 105 V | 5 nA | 0.020% + 15 nA | 100 pA | 100 pA | <100 µV | 0.020% + 10 nA |
| 1.000000 mA | 105 V | 50 nA | 0.020% + 150 nA | 1 nA | 1 nA | <100 µV | 0.020% + 100 nA |
| 10.00000 mA | 105 V | 500 nA | 0.020% + 1.5 μA | 10 nA | 10 nA | <100 µV | 0.020% + 1 μA |
| 100.0000 mA | 105 V | 5 μA | 0.020% + 15 μA | 100 nA | 100 nA | <100 µV | 0.020% + 10 μA |
| 1.000000 A | 105 V | 50 μA | 0.050% + 750 μA | 5 μA | 1 μA | <100 µV | 0.050% + 500 μA |
| 4.000000 A | 21 V | 250 μA | 0.100% + 3 mA | 25 μA | 1 μA | <100 µV | 0.100% + 2.5 mA |
| 5.000000 A | 10.5 V | 250 μA | 0.100% + 3 mA | 25 μA | 1 μA | <100 µV | 0.100% + 2.5 mA |
| 7.000000 A | 7.35 V | 500 μA | 0.150% + 6 mA | 125 μA | 1 μA | <100 µV | 0.150% + 5 mA |