Capnography HP ETCO2 Sensor Mainstream For MP5 MP2 X2 M2501A

Price Negotiable
MOQ: 1pcs
Delivery Time: About one week
Brand: Med Accessories
Product Description

M2501A Mainstream Capnography HP EtCO2 Sensor for MP5 MP2 X2     

Compatibility
Manufacturer Model
HP Compatible M3002A MMS X2, M8102A MP2, M8105A MP5
Project Specifications
Principle of Operation Non-dispersive infrared(NDIR) single beam optics, dual wavelength, no moving parts.
Initialization Time Capnogram displayed in less than 3 seconds,full specifications within 3 minutes(At 25°C).
CO2 Measurement Range 0 - 114 mmHg; 0 - 15 %; 0 - 15.2 kPa.
CO2 Rise Time Rise Time Less than 80 ms - Adult Reusable or Single-Patient-Use Airway Adapter Less than 80 ms - Infant Reusable
or Single-Patient-Use Airway Adapter.
CO2 Resolution 0. 1mmHg ; 0 to 38 mmHg.; 0.25 mmHg ; 39 to 114mmHg.
CO2 Accuracy 0 - 40mmHg ±2 mmHg.; 41 - 76mmHg ±5% of reading.; 77 - 114 mmHg ±8% of reading.
Above 80 breaths per minute ±12% of reading. * Note: Gas emperature at 25°C.
CO2 Stability Short term drift: Drift over four hours shall not exceed 1 mmHg maximum.
Long term drift: Accuracy specification will be maintained over a 120-hour period.
Sampling Frequency 100Hz.
Respiration Rate Range 3 to 150 breaths per minute.
Respiration Rate Accuracy ±1 breath.
Calibration No routine user calibration required. An airway adapter zero is required when changing to a different style of airway
adapter.
Airway Adapters Single-patient-use or reusable, < 5 cc dead space (adult), < 1 cc dead space infant.
Temperature Operating: 0 to 40°C; Storage: -40 to 70°C; Transport: -40 to 70°C.
Humidity Operating: 10 to 90% RH, non-condensing; Storage: 10 to 90% RH, non-condensing.;
Transport: 10 to 90% RH, non-condensing.
Atmospheric Pressure Operating: 400-800 mmHg; Storage: 400-800 mmHg; Transport: 400-800 mmHg.
Water Resistance IPX4 - Splash-proof - Sensor only.
Voltage Requirements 5VDC ± 5 %.
Power Rating <1.0W Typical; <1.8 W Warm-up.
Sensor Weight <85g.
Sensor Size Deep*Wide*High:48mm * 24.3mm * 35.5mm.
Cable Length 2.4m.

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Med Accessories Technology Dongguan Co., Ltd.
Location No.210,211,212,213,214,Building 1,No.12 Industry Western Road,SongShanHu,Dongguan,Guangdong,China.523808
Contact Person Carina

Request A Quote

Please check your email address.
Your message must be at least 20 characters.