SICK Analyzer Module DEFOR for Series GMS800 Cl2, H2S, NO, NO2, NH3, SO2 COS CS2 and further gases

Price Negotiable
Price: Negotiable
MOQ: 1
Delivery Time: 15-20days
Brand: SICK
Product Description

Measuring system
Measuring principle
The Analyzer module DEFOR uses the fact that some gases have a specific absorption
characteristic in the ultraviolet light range. For this purpose, UV light is radiated through the
sample gas. The concentration of a gas component in a gas mixture can be determined
through suitable selection of the light wavelength and selective absorption measurement.
The Analyzer module DEFOR can, in this way, analyze the concentration of the gases Cl2,
H2S, NO, NO2, NH3, SO2 and further gases.
The Analyzer module DEFOR can simultaneously measure up to 3 gas components.
Measuring method
● The Analyzer module DEFOR uses interference filter correlation (IFC) for most gas com-
ponents. Measuring and reference radiation are created alternately by swiveling two dif-
ferent interference filters into the beam path (filter wheel [2a]).
● Gas filter correlation (UVRAS) is used for NO. This method creates the reference radia-
tion by swiveling a gas filter filled with the relevant gas into the beam path (filter wheel
[2b]).
● Both measuring methods are combined for joint measurement of NO and other gases
(both filter wheels [2a]+[2b] are used).
● The physical state of the measuring system is captured and compensated using a refer-
ence beam path through a reference cuvette.
Analyzer module layout
–Beam source is a special UV gas discharge lamp that emits both broad-
band as well as NO-specific radiation components.
–Thermostat-controlled filter wheel unit keeps the optic filters at a constant tempera-
ture. This minimizes the influence of external temperature changes.
–Lenses , beam splitter and mirror direct the beam path.
–Sample gas flows through sample cuvette. Reference cuvette is either filled with
a neutral gas or has span gas flowing through (option).
–Detectors capture the radiation intensity created with the various filters.
–Measurement signals are amplified electronically and digitally evaluated. The structure
symmetry compensates proportional and symmetric signal drifts in the best possible
way.
–The measuring system can be fitted with an adjustment unit

Similar Products

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company WUHU WANYI INDUSTRY CO.,LTD
Location No7 .Baizhuangkuajie zhanghe road, Yijiang zone, Wuhu City,Anhui Province,China 241000
Contact Person Yolanda

Request A Quote

Please check your email address.
Your message must be at least 20 characters.