YC1893D Three-Phase Meter Test Bench with 300V Voltage Output and 120A Current Range for Electricity Meter Calibration
Price:
negotiation
MOQ:
1 Piece
Delivery Time:
4-8weeks
Brand:
GENY
Product Description
The YC1893D Professional Three-Phase Meter Test Bench is a high-precision calibration system designed for electricity meter verification. This fully automatic test bench provides reliable and accurate calibration for single-phase and three-phase electric energy meters, serving laboratories, utility companies, and meter manufacturers with exceptional performance.
System Overview
This versatile smart meter testing system accommodates various meter types including mechanical, electronic-mechanical, and electronic smart meters. It supports automated, semi-automatic, and manual operation modes, ensuring flexible and precise electricity meter calibration for production, laboratory, or industrial environments.
System Components
- Three-phase power source for precise voltage and current output
- Three-phase reference standard meter for high-accuracy calibration
- Meter suspension racks compatible with single-phase and three-phase meters
- Smart meter function testing system with optional features
- PC and software for automated control, data acquisition, and test management
- Complete auxiliary components for comprehensive electricity meter testing
Standards Compliance
The YC1893D meter test system is fully compliant with international standards including IEC 60736, IEC 62052-11, IEC 62053 series (accuracy classes 0.2, 0.5, 1.0), IEC 62056, EN 50470 series, and IEC 61010 for electrical safety. These certifications ensure IEC-compliant meter testing for both single-phase and three-phase meters, including smart meters.
Technical Specifications
Power Source
| Parameter | Value |
|---|---|
| Test voltage output (Phase-Neutral) | 3x(24 - 300 V) |
| Power of the voltage output | 3×1000VA |
| Resolution | Better than 0.01% of full scale value of range |
| Setting accuracy | Better than 0.05% at the final range value |
| Stability | Better than 0.005%/h (integration time 150 s) |
| Load regulation from 0-max. load | < 0.01% |
| Distortion factor | < 0.3% for linear resistance load |
| Harmonics | 2nd - 21st free programmable |
Current Output
| Parameter | Value |
|---|---|
| Test current output | 1 mA - 120 A |
| Power of the current output | 3×2000VA |
| Resolution | Better than 0.01% of full scale value of range |
| Setting accuracy | Better than 0.05% at the final range value |
| Stability | Better than 0.005%/h (integration time 150 s) |
| Load regulation from 0-max. load | < 0.01% |
| Distortion factor | < 0.3% for linear resistance load |
| Harmonics | 2nd - 21st free programmable |
Phase Angle & Frequency
| Parameter | Value |
|---|---|
| Phase angle range | 0 - 360° |
| Phase angle resolution | 0.01° |
| Phase angle setting accuracy | 0.1° |
| Frequency range | 45 Hz - 65 Hz |
| Frequency resolution | 0.01 Hz |
Related Videos
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.
Company
Guangzhou GENY Electric Co., Ltd
Location
ROOM 107, 1ST FLOOR, No. 602 GUANGSHAN ER ROAD, TIANHE DISTRICT, GUANGZHOU, CHINA POSTCODE: 510663
Contact Person
John Shen