Advanced Energy Meter Test Bench for Utility and Laboratory Use

Price Negotiable
Price: Negotiation
MOQ: 1 pcs
Delivery Time: 30 days after order
Brand: GENY
Product Description
Advanced Energy Meter Test Bench for Utility and Laboratory Use
Standard Functions
  • Accurate measurement of active power, reactive power, active energy, and reactive energy
  • Start-up and creep tests for electricity meters
  • Automatic mark function for fast start-up and creeping verification
  • Measurement of basic error, average error, standard deviation, and 24-hour variation
  • Voltage influence, frequency influence, harmonic influence (2nd-21st), and phase-sequence influence tests
  • Simultaneous current measurement across all meter positions
  • Testing without disconnecting the meter's voltage and current circuits
  • Built-in current open-circuit protection and voltage short-circuit protection
  • High-efficiency optical scanning head for pulse acquisition
  • Supports both motorized and manual operating modes
  • Software-based automatic error calculation, data sorting, data storage, and online output
  • Data management with customizable report printing
  • Barcode input supported
  • Simple, durable, and reliable mechanical structure
  • Flexible configurations from 3 to 60 meter positions
Special Functions
  • Forward and reverse active energy error tests
  • Forward and reverse reactive energy error tests
  • Compatible with SW-5D precision time-base source for daily timing error measurement
  • Communication test and broadcast time synchronization for smart meters
  • Meter constant verification (pulse counting / timing / start-stop energy methods)
  • Time-period switching, demand interval, and demand indication error tests
  • Register indication and combined register error tests
  • Power consumption test (optional)
  • Inclination influence test (optional)
  • Prepaid energy meter test functions (optional)
Specifications
Main Technical Indicators Parameter Value
Power Supply AC220V±10%, 50Hz±2Hz
Ambient Temperature -10°C ... +40°C
Ambient Humidity 35% ... 95%
Output Voltage Range 24V ... 288V
Output Voltage Power 1000VA (Max)
Voltage Resolution 0.01%
Voltage Stability 0.02%/120s
Load Change Rate 0.01%
Voltage Distortion ≤0.3% (Full linear load range)
Output Current Range 1mA ... 120A (Up to 230A Max)
Output Current Power 2000VA (Max)
Current Resolution 0.01%
Current Stability 0.02%/120s
Load Change Rate 0.01%
Current Distortion ≤0.3% (Full linear load range)
Continuous Current 100A
Phase Adjustment Range 0° ... 360°
Phase Resolution 0.01°
Frequency Adjustment Range 45Hz ... 65Hz
Frequency Resolution 0.01Hz
Power Output Stability 0.02%/120s
Advanced Energy Meter Test Bench for utility and laboratory applications

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.