Product Description

What it is

The Liquiphant FTL64 is a point-level switch (limit level detector) manufactured by Endress+Hauser. It uses a vibronic measurement principle (a vibrating fork) to detect the presence or absence of liquid at a given point (i.e., a “limit” or “alarm” level) in a tank, vessel or pipeline.
It is designed specifically for high‐temperature and demanding industrial applications (hot liquids, pumpable fluids, high pressures).

 

How it works

  • The device contains a “fork” or “tuning-fork” (vibrating element) that is excited at a known frequency. When immersed in the liquid, the vibration is damped; when the liquid is absent, the damping changes. That change is detected and signals the switch. (General vibronic principle)
  • Because it’s a point detector, it gives a “liquid present / liquid not present” signal, rather than a continuous level measurement.
  • The FTL64 is offered with digital communication options, diagnostics (Heartbeat Technology by Endress+Hauser) and supports modern asset management and smart commissioning via e.g., Bluetooth/SmartBlue app.

 

Key Specifications & Features

Here are some of the standout technical specifications and features (note: always check the exact model variant for your application):

  • Process temperature range: from –60 °C up to +280 °C. In some versions, up to +300 °C (for limited period) and with PFA coating up to ~230 °C.
  • Process pressure range: from vacuum up to ~100 bar (depending on variant) for the normal version; PFA coated versions have lower pressure limit (e.g., up to ~40 bar) in some cases.
  • Medium density (min.): typical minimum ~0.5 g/cm³ (optionally 0.4 g/cm³) for the liquid to be detected.
  • Viscosity capability: Up to ~10,000 mm²/s (cSt) — suitable for “pumpable” liquids/fluids.
  • Materials / coatings: Choice of wetted materials (e.g., 316L stainless steel) and optional PFA coating for aggressive media.
  • Process connections: Threads, flanges, hygienic connections (Tri-Clamp) for varied installation requirements.
  • Safety & approvals: Developed per IEC 61508; rated for SIL2/3 in some variants; certifications for explosion protection, marine approvals, over-fill protection (e.g., WHG) included.
  • Diagnostics & maintenance: Heartbeat technology enables continuous diagnostics; guided proof testing via app or test pin/button from outside — simplifying safety loop testing.

 

Where / Why You Would Use It

Typical application scenarios for the FTL64 include:

  • High‐temperature liquids in industrial processes (e.g., chemical, petrochemical, power generation) where pumps, pipelines or vessels need level detection in hot fluid conditions.
  • Over-fill protection of tanks storing hot liquids. The robust design suits extreme temperature conditions.
  • Dry‐run protection for pumps in hot liquid systems (i.e., detecting when liquid level drops below pump suction).
  • Applications where alternatives (float switches, optical sensors) fail due to temperature, viscosity, media aggressiveness, deposits, bubbles or other difficulties. The vibronic switch is less sensitive to such interferences.

In short: when you need a reliable level switch under harsh conditions (high T, possibly high pressure, viscous or aggressive media) — this product fits that niche.

 

Benefits / Key Advantages

  • High reliability even in tough environments: high T, high P, viscous fluids.
  • Simple yet robust limit detection: less maintenance than some more complex level measurement technologies in harsh conditions.
  • Strong safety features and diagnostics built-in: ideal for critical installations requiring SIL/over-fill protection.
  • Flexibility of installation: multiple connection types, materials, coatings make it adaptable.
  • Modern commissioning/maintenance: Bluetooth/SmartBlue app, heartbeat diagnostics simplify asset management and reduce downtime.

 

Considerations / Things to Check

  • Ensure the version you select supports your maximum process temperature – e.g., if you will exceed +230 °C, ensure the variant supports up to +280 °C (or the short term +300 °C).
  • Check the pressure rating for your specific application and verify the process connection’s rating as this may define the maximum safe pressure.
  • Minimum medium density must be met: if you have extremely low‐density liquids, check whether the 0.4 g/cm³ option is needed.
  • For highly aggressive media, consider the PFA-coated variant — but note that for PFA coated version the pressure/temperature limits may differ (lower) so review carefully.
  • Even though it is robust, ensure good installation practice: correct immersion, location free of unwanted turbulence or splashing, correct orientation, and maintenance access for proof testing.
  • Check compatibility with your control and safety system: output signals, diagnostics, communication (HART, Bluetooth etc) must integrate with your system.
  • Consider the lifecycle & document the proof-testing regime: for safety critical usage (SIL, over-fill protection) you’ll want documented testing/maintenance procedures.

 

 

Category

Description

Product Type

Vibronic (tuning fork) point level switch for liquids

Manufacturer

Endress+Hauser

Measurement Principle

Vibrating fork (frequency damping changes when immersed in liquid)

Primary Function

Reliable detection of liquid presence/absence (limit detection)

Typical Applications

Overfill protection, dry‐run protection, level limit detection in tanks and pipelines

Process Temperature Range

–60 °C … +280 °C (up to +300 °C short term)

Process Pressure Range

Vacuum … 100 bar (depending on version)

Medium Density Range

≥ 0.5 g/cm³ (optionally 0.4 g/cm³ for low‐density liquids)

Viscosity Range

Up to 10,000 mm²/s (cSt)

Wetted Materials

Stainless steel (316L / 1.4404), optional PFA coating for aggressive media

Process Connections

Threads, flanges, Tri-Clamp/hygienic connections

Electrical Output / Communication

2-wire (AC/DC), relay, transistor, PFM, or digital (HART, Bluetooth)

Integrated Diagnostics

Heartbeat Technology – continuous monitoring and guided proof testing

Safety & Approvals

SIL 2/3 (IEC 61508), WHG overfill protection, Ex approvals (ATEX, FM, CSA, etc.)

Key Features

Robust at high T & P, maintenance-free operation, unaffected by foam/bubbles, simple proof testing

Installation Orientation

Horizontal or vertical (as per application)

Maintenance

Virtually none; proof testing via Heartbeat or external test button

Industry Sectors

Chemical, oil & gas, power generation, food & beverage, water treatment

 

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Xi'an Xiyi Central Control Technology Co., Ltd.
Location Room 2204, Unit 1, No. 91, Taoyuan South Road, Xi'an, Shaanxi, China
Contact Person Lee

Request A Quote

Please check your email address.
Your message must be at least 20 characters.