Endress Hauser Digital conductivity sensorMemosens  CLS15E-GRHB10B

Field of application

 

Memosens CLS15E is perfectly suited for conductivity measurement in applications with low measuring ranges such as boiler feedwater and chip cleaning. The contacting conductivity sensor performs reliably and accurately even in hazardous areas. CLS15E features Memosens 2.0 digital technology, offering extended storage of calibration, adjustment and process data. It facilitates predictive maintenance and provides the perfect basis for IIoT services.

 

Memosens CLS15E measures conductivity in pure and ultrapure water applications for the monitoring and control of:

  • Boiler feedwater

  • Ion exchangers

  • Reverse osmosis

  • Cooling water

  • Distillation

  • Chip cleaning

The digital sensor can be used in power plants for:

  • Degassed conductivity

  • pH measurement with differential conductivity

With its international explosion protection approvals, such as ATEX and FM, the sensor is suitable for hazardous area use.

 

Potentiometric

 

Measuring principle

Conductive

 

Application

Measurement in pure and ultrapure water range

Monitoring of ion exchangers

Reverse osmosis

Distillation and chip cleaning

 

Characteristic

Digital 2-electrode conductivity sensor

 

Measuring range

k=0,01: 0.04 to 20 µS/cm

k=0,1: 0.10 to 200 µS/cm

 

Measuring principle

Conductive conductivity cell with electropolished stainless steel electrodes

 

Design

2-electrode conductivity cell with coaxially arranged electrodes, electropolished

 

Material

Electrode: stainless steel 1.4435

Electrode shaft: PES

 

Dimension

Electrode diameter: 16 mm (0.63 inch)

Electrode length: appr. 55 mm (2.17 inch)

 

Process temperature

Threaded with fixed cable:

-20 to 100 °C (-4 to 212 °F)

Threaded with plug-in head:

-20 to 120 °C (-4 to 248 °F)

Sterilization: max. 140 °C (284 °F) for 30 minutes

 

Process pressure

13 bar at 20 °C (188 psi at 68 °F) absolute

2 bar at 120 °C (14 psi at 248 °F) absolute

 

Temperature sensor

Pt1000

 

Ex certification

ATEX, NEPSI, CSA, IECEx, INMETRO, EAC Ex

 

 

Ingress protection

IP68

 

Additional certifications

Calibration certification of the cell constante and temperature, material certification 3.1

 

Connection Process: 1/2" and 3/4" NPT, Clamp 1 1/2"
   
  Sensor connection: Inductive, digital connection head with Memosens 2.0 technology

 

 

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Achievers Automation Limited
Location RM 509, 5/F, THE CLOUD, 111,TUNG CHAU STREET, TAI KOKTSUI, KOWLOON,HONG KONG
Contact Person Joanna Yang

Request A Quote

Please check your email address.
Your message must be at least 20 characters.