AC220V 50Hz DSC Differential Scanning Calorimetry Instrument RT 600℃
DSC-200 Differential scanning calorimeter (DSC)
Product introduction
DSC-200 is one of the most accurate DSC series products launched by the company. The sensor is made of imported material E-pair, which has high accuracy, high sensitivity and good repeatability. The signal acquisition circuit has shielding protection, strong anti-interference, and extremely high baseline stability and repeatability.
DSC-200 differential scanning calorimeter can be used to test glass transition temperature, phase transition temperature, melting point, enthalpy value, curing temperature, product stability, oxidation induction period, etc. It is competent in the research of pipes, polymers, chemicals, food, medical treatment and many other fields, and its products serve universities, enterprises, third-party measurement and quality inspection units, with a wide range of applications to meet the testing needs of various industries.
Main Features
- Industrial level of the 7-inch touch screen, display rich information
- New furnace body structure, better baseline, higher accuracy. Heating adopts indirect conduction mode, high uniformity and stability, reduce pulse radiation, better than the traditional heating mode.
- USB communication interface, strong versatility, reliable communication is not interrupted, support self-recovery connection function.
- Automatic switching of two atmosphere flow rates, fast switching speed, short stability time. At the same time, increase the protective gas input.
- Software is simple and easy to operate.
- Ultra-high sensitivity, precision, 0.001 mw, 0℃.001
- Exquisite technical indicators, superior performance, high-cost performance, widely used
Technical parameters
|
MODEL |
DSC-200 |
|
Temperature range |
Room temperature of ~600℃ |
|
Temperature resolution |
0.001℃ |
|
Temperature fluctuation |
±0.001℃ |
|
Temperature repeatability |
±0.01℃ |
|
Heating rate |
0.1~100℃/min |
|
The constant temperature time |
Was programmed for 24h |
|
Temperature control mode |
Heating up, constant temperature, cooling (fully automatic program control) |
|
DSC range |
0~±600mW |
|
DSC, resolution |
0.01uW |
|
DSC, sensitivity |
0.001mW |
|
Working power supply |
AC220V / 50Hz or customized |
|
The atmosphere controls the gas |
Nitrogen, oxygen (programmed / automatic switching) |
|
Gas-flow rate |
0-300mL/min |
|
Gas pressure |
0.2MPa |
|
Display mode |
24bit color, 7-inch LCD touchscreen display |
|
Data interface |
Standard USB interface |
|
Parameter standard |
Equipped with a standard material (indium, tin), the user can correct the temperature by himself |
|
Optional Cooling device |
Air cooling device (optional semiconductor -40~550℃, liquid nitrogen refrigeration -150~550℃) |
|
Thermocouple |
with multiple sets of thermocouples, temperature of a group of measured samples,One group to measure furnace temperature,One group measures the internal ambient temperature |
Test map: Glass transition temperature

Test map: melting point and heat enthalpy

Lift temperature scan

Get in Touch