Product Description

 

Introduction

The FN-IV Falling Number Tester (with printer) is manufactured by our company in accordance with the new national standard GB/T 10361-2024 "Wheat, Rye and their Flours and Durum Wheat and its Semolina - Determination of Falling Number by the Hagberg-Perten Method". It is a specialized instrument for measuring the α-amylase activity in grains, accurately determining the degree of germination damage in grains. It is suitable for testing grains, especially wheat and wheat flour, and is an essential instrument for quality testing in grain storage, flour processing, food processing, universities, and research institutions. This instrument is a dual-tube falling number tester, consisting of a testing unit (water bath, stirring system), a control unit display screen, and a printer. The display screen shows the falling number. The control unit uses a high-performance, low-power single-chip microcomputer, which accurately records the two sets of falling numbers for each measurement and the actual date of each measurement. It also uses a miniature printer with a standard Chinese character library and front paper loading, which automatically prints the sample number, falling number, and date.

 

AFN-IV Falling Number Tester High Accuracy Grains Germination Damage Degree Test Machine Grains Analyzer

Technical Specification

Rubber stopper

Matches the viscosity tube

Heating tube power

600 W

Electronic timer

Timing function

Stirring rod weight

(25.00 ± 0.05) g

Precision viscosity tube

Made of special glass

Inner diameter

21.00 mm ± 0.02 mm

Outer diameter

23.80 mm ± 0.25 mm

Inner height

220.0 mm ± 0.3 mm

Water bath device

Composed of an integrated heating unit, cooling system, and water level indicator

Repeatability

The difference between two independent test results obtained in the same laboratory, by the same operator, using the same equipment and the same test method, on the same test object within a short period of time, shall not exceed 8%

Measurement error

≤ 5%

Working voltage

AC 220 V ± 10%, 50 Hz

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company DONGGUAN LONROY EQUIPMENT CO LTD
Location A31, 11 Floor, No 1 Building, Minjian Jinrong Building, No 111 Guantai Road, Sanyuanli Community, Nancheng Street, Dongguan, Guangdong, China
Contact Person Kaitlyn Wang

Request A Quote

Please check your email address.
Your message must be at least 20 characters.