TF-30D201 Zirconia Oxygen Analyzer for Industrial Semiconductor Atmosphere N2 Furnace Air Separation Unit 741934634 971351824
Price:
Inquire
MOQ:
NO MOQ
Delivery Time:
Inquire
Brand:
EVOzuRun
Product Description
| Specification | Details |
| Model | TF-21D201 (TF-21 Sensor + DTF-201 Receiver) |
| Measuring Principle | Zirconia type concentration cell method |
| Measuring Range | Minimum: 0–10 ppm; Maximum: 0–25%. Any four ranges selectable with a fixed zero point. Output range configurable from 0–10 ppm to 0–100%. |
| Calibration | Air-point calibration plus up to four additional points. Calibration available for each measurement range. |
| Analog Output | DC 4–20 mA, non-isolated output, load resistance ≤600 Ω. Selectable DC 0–1 V / DC 0–5 V / DC 0–10 V, non-isolated output. |
| Sample & Calibration Gas Flow Rate | 0.5 L/min |
| Accuracy — Repeatability | ±1% FS for ranges below 0–1%; ±0.5% FS for 0–1% range and above. |
| Accuracy — Linearity | ±3% FS for ranges below 0–100 ppm; ±2% FS for ranges below 0–1%; ±1% FS for 0–1% range and above. |
| Response Time | ≤30 seconds for ranges below 0–1%; ≤10 seconds for 0–1% range and above. Based on 90% response when switching calibration gas. |
| Drift | ≤±2% FS per week |
| Power Supply | AC version: AC 100–240 V ±10%, 50/60 Hz, 50 VA. DC version: DC 24 V ±10%, 50 W. |
| Contact Input | Range switching via voltage-free contact |
| Contact Output | Open collector output, DC 25 V max., 50 mA. Four outputs. Selectable functions: Analyzer fault / Range echo-back 1 / Range echo-back 2 / Range echo-back 3 / Ready / High concentration alarm / Low concentration alarm. |
| Range Switching | Local / Remote / Automatic range switching |
| Output Hold Function | Selectable: No hold / User-defined value / Value recorded 5 seconds before fault occurrence. Output held during warm-up and analyzer fault when hold is enabled. |
| High/Low Concentration Alarm | High and low alarms independently configurable from 0.0–9,990.0 ppm or 0.0–100.0%. |
| RS-232C Communication Function | Transmits measurement status, error code, measurement range, and measured value. Communication cable: D-sub 9-pin, straight cable. |
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.
Company
Guangzhou Zhuyou Auto Parts Co., Ltd.
Location
Room 418, Haoquan Building, Jingtai Street ,Guangzhou,Guangdong,China
Contact Person
Suxi