The AG-GC07 gas chromatograph uses high-temperature heateddual-column parallel backflushing chromatography separation technology.
lt can automatically measure and analyze methane and total hydrocarbons content. The non-methane totalhydrocarbons (NMHC) content is calculated by the difference between total hydrocarbons and methane. Thissignificantly shortens the analysis cycle. Additionally, the high-temperature heating technology, designed forhigh-boiling non-methane total hydrocarbons, greatly reduces the chromatographic peak broadening ofhigh-boiling NMHCs, allowing for accurate measurement of NMHCs, even in the presence of high concentra-tions of non-methane total hvdrocarbons.
>>Feature
1. Uses mass flow meter control to ensure measurement precision and accuracy.
2. Passivated tubing treatment to reduce gas adsorption and residue.
3. Flameout automatic gas source cutoff for enhanced safety.
4. Capable of simultaneous monitoring of non-methane total hydrocarbons, benzene series, and othercharacteristic factors.
>>Technical parameters
| Dimensions | 19'' standard chassis, 5U height |
| Power supply | 220VAC,50Hz |
| Operating Power | <500VA(Startup <1000VA) |
| Operating Temperature | (5-35)ºC |
| Operating Humidity | (20-90)%RH |
| Detector | High-sensitivity FID detector |
| FID Detection Limits | Methane 0.1mg/m3; Non-methane total hydrocarbons (NMTH) 0.05mg/m3 (as propane) |
| Range | Methane 0.1~100mg/m3; NMTH 0.1~100mg/m3 as propane, customizable range |
| Chromatographic Column | 1m Porapak Q packed column |
| Analysis Cycle | 2min |
| Sample Gas Temperature | Ambient to~150ºC |
| Sample Flow Rate | ≥0.5L/min |
| Gas Supply Requirements | Hydrocarbon-free air:0.4MPa; Hydrogen:0.2MPa; Nitrogen:0.4MPa; Calibration gas:CH4/C3H8 |
| Repeatability | <2% |
| Auxiliary Equipment | Built-in industrial control computer |
| Analog I/O | 485 |