UL1581 ASTM D2436 Air Ventilation Aging Testing Chamber

Price Negotiable
Price: Negotiable
MOQ: 1 set
Delivery Time: 7~10 days after the order confirmed
Brand: Zhongli
Product Description
UL1581 ASTM D2436 Air Ventilation Aging Testing Chamber
Professional testing chamber designed for air ventilation aging tests according to UL1581 and ASTM D2436 standards, ensuring precise environmental simulation for material testing applications.
Performance Specifications
Parameter Specification
Power Supply AC380V
Power 6KW
Temperature Range Ambient temperature +25℃~300℃
Control Accuracy ±0.5℃
Resolution Accuracy 0.1℃
Distribution Error ±2.5℃
Heating Time From ambient temperature to 300℃, approximately 60 minutes
Machine Structure & Materials
  • Inner chamber dimensions: 45cm (depth) * 50cm (width) * 50cm (height)
  • Outer dimensions approximately: 1040mm (length) * 650mm (width) * 1275mm (height)
  • Inner chamber material: SUS#304 stainless steel, outer case: baked paint
  • Insulation material: High-density rock wool
  • Heater: Spiral tube heater set: AC380V 6KW
Air Exchange Rate Test Procedure
Comprehensive testing methodology for determining air exchange performance in aging test chambers:
  • Seal all ventilation holes, doors, thermometer holes, and gaps around the blower using pressure-sensitive tape
  • Connect a standard watt-hour meter to the aging chamber's power system
  • Control working chamber temperature to 80±2℃ above ambient temperature
  • Measure electrical energy consumption after temperature stabilization for 2 hours
  • Remove seals and repeat measurement with air exchange port open
  • Calculate air exchange rate using standardized formula
Air Exchange Rate Calculation
N = 3590 • (P2 - P1) / V • d • △T
Where:
  • N = Air exchange rate (times/hour)
  • P1 = Average electrical power consumed when sealed (W)
  • P2 = Average electrical power consumed when open (W)
  • △T = Difference between chamber temperature and ambient temperature (℃)
  • V = Working chamber volume (L)
  • d = Density of air at surrounding ambient temperature (g/L)
Forced convection types: Up to 100 times/hour or 100-200 times/hour (adjustable)
UL1581 ASTM D2436 Air Ventilation Aging Testing Chamber equipment

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Dongguan Zhongli Instrument Technology Co., Ltd.
Location No.62, Xinhe Management of Area, Wanjiang District, Dongguan city, Guangdong province ,China
Contact Person Fiona Zhong

Request A Quote

Please check your email address.
Your message must be at least 20 characters.