7.5mm 15mm Probe TPS 130c Thermal Conductivity Tester

Price Negotiable
Price: Negotiable
MOQ: 1 set
Delivery Time: 5-8 work days
Brand: BAXIT
Product Description

 

operation video
 

 

Testing object:

Suitable for metal, ceramic, alloy, ore, polymer, composite, paper, fabric, foam (surface-level insulation, sheet), mineral wool, cement wall, glass reinforced composite board CRC, cement polystyrene board, sandwich Concrete, FRP panel composite panels, paper honeycomb panels, colloids, liquids, powders, granules and paste solids, etc., have a wide range of test objects.

Working principle:

Transient planar heat source technology (TPS) is a new method for measuring thermal conductivity, developed by Professor Silas Gustafsson of the University of Technology, Chalmer, Sweden, based on the hotline method. Its principle of determining the thermal properties of a material is based on the transient temperature response produced by a step- heated disk-shaped heat source in an infinite medium. A flat probe is made of a thermally resistive material, and serves as both a heat source and a temperature sensor. The thermal resistance coefficient of the alloy has a linear relationship with the relationship between temperature and resistance. That is, the heat loss can be known by understanding the change of the resistance, thereby reflecting the thermal conductivity of the sample.The continuous double-spiral structure sheet formed by etching after gold etching has a double-layer insulating protective layer and a thin thickness, which makes the probe have a certain mechanical strength and maintain electrical insulation from the sample. During the test, the probe was placed in the middle of the sample for testing. When the current passes through the probe, a certain temperature rise occurs, and the generated heat is simultaneously diffused to the samples on both sides of the probe. The speed of thermal diffusion depends on the heat transfer characteristics of the material. By recording the temperature and the response time of the probe,The thermal conductivity can be directly obtained from the mathematical model.

 

Technical parameters:

Test range 0.005—300W/(m*K)
Measuring temperature range room temperature - 130 ° C
Probe diameter No. 1 probe 7.5mm; No. 2 probe 15mm
Accuracy ±3%
Repeatability error ≤3%
Measurement time 5~160 seconds
Power supply AC 220V
Machine power <500w
Sample temperature rise <15 ° C
Test sample power P No. 1 probe power 0 < P < 1w; No. 2 probe power 0 < P < 14w
Sample specifications Single sample (15*15*3.75mm) measured by the first probe; single sample (30*30*7.5mm) measured by the second probe.
Note: The No. 1 probe measures a thin material with a low thickness. The sample is superimposed if the surface of the sample is smooth and flat and sticky.

 

methods Transient plane heat source method laser method hot-wire technique guarded plate method
  not steady state method not steady state method not steady state method steady state method
measuring physical properties Thermal conductivity and thermal diffusivity are obtained directly. The thermal diffusivity and specific heat are obtained directly, and the thermal conductivity is calculated by the input sample density value. Thermal conductivity is obtained directly Thermal conductivity is obtained directly
application range solid, liquid, powder, paste, colloid, particle solid Solid, liquid solid
sample preparation No special requirements, simple sample preparation. Sample preparation is multifarious. Sample preparation is simple and has specific requirements. Larger sample size.
accuracy of measurement ±3%,better±0.5% better±10% better±5% better±3%
physical model Planar heat source contact measurement, as long as the finite surface contact is good. The heat source is contactless. The linear heat source must be in good contact with the linear model. Heat source contact type, need good surface contact.

Thermal conductivity range

[w/(m*k)]

0.005-300 10-500 0.005-10 0.005-5
measuring time 5-160S few minutes few minutes few hours

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Shanghai Glomro Industrial Co., Ltd.
Location No.5616th, Caoan Road, Jiading District, Shanghai
Contact Person Clarke Wang

Request A Quote

Please check your email address.
Your message must be at least 20 characters.