China World Technology Medical Equipment Service Group
                                                                                                           
Verified Supplier
11 Years
Since 2015
Menu

PN 8422711 Savina Ventilator Power Supply EOE12060720

Price Negotiable
Price: Negotiable
MOQ: 1pcs
Delivery Time: 3-5 working days
Brand: Dräger
Product Description

PN 8422711 Dräger Savina Ventilator Power Supply EOE12060720

 

 

Power supply

 

Savina runs on mains power or an internal battery. Savina can be powered by an external battery or built-in DC power supply. An external battery

 

is also used as a power source during patient transport.

 

Savina is designed for connection to the hospital's mains power supply.

 

 

WARNING

 

Risk of electric shock and device malfunction If the device is connected to a socket with the wrong mains voltage or without a protective earth

 

connection, the user is at risk and the device could be damaged. protective earth connector

 

NOTE

 

In operation, the power socket used must be readily accessible.

 

 

 

Power supply from on-board power supply or batteries

 

WARNING

 

Risk of explosion Electrolytic gas can occur when the batteries are charging. In a sufficient concentration, this can cause an explosion. The device

 

must always be placed in a wellventilated area when connected to mains power.

 

 

Supply from a DC on-board power supply

 

Savina can be powered by a DC on-board power supply

 

 

Connecting the power supply

 

Prerequisites: – Installation and connection by service personnel only. – The poles of the DC vehicle power supply must be protected against

 

contact

 

Warn

Risk of patient injury If the poles of the DC vehicle power supply are not protected against contact, excessive patient leakage currents may result.

 

Prevent poles from touching.

 

Warn

RISK OF EQUIPMENT DAMAGE - Do not connect mains powered equipment to the external battery connector. – Only external batteries included

 

in the accessories list can be connected. – Only use the connecting cables listed in the accessories list.

 

 

The plug on the device side of the connecting cable is coded. This allows Savina to detect if an external battery or DC vehicle power supply is connected.

 

 

Specification

Input 100-240V~/3,4-1.3A;50-60Hz
Ext 24V-/13.5A
Int 24V-/11.5A
Output 5V-/4.5A;15V-/1.5A;-15V-/1.3A 24V-/1.2A;48V-/2.4A
Charge 0-32V-/45W (27V-/1.66A)

 

 

Failure of the power supply

If the power supply fails, Savina generates a power supply failure alarm. The ventilation settings and the alarm limits remain saved even in the

 

event of a power supply failure

 

  • Restore the power supply immediately
  • Disconnect patient from the device and continue ventilation without delay using another independent ventilator.

Thank you so much for your attention.If you're interested in PN 8422711 Dräger Savina Ventilator Power Supply EOE12060720 ,feel free to contact us .Thank you .

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.