 Parameter Panel Board 453564200151 of  PHiliphh VS2+ Patient Monitor is made of plastic ,which includes NBP and Spo2 connector .The panel connector is to connect NBP and Spo2 accessories to make NBP and Spo2 function 

 

 

NBP Test

 

The NBP test examines the performance of noninvasive blood pressure measurements. When checking the NBP module, perform the following steps:

 

  • NBP Accuracy NBP Calibration Procedure (if required)

 

  • NBP Pneumatic Leak Test

 

To perform this test, you need the following tests:

 

  • Reference pressure gauge (including hand pumps and valves)

 

  • Reading expansion chamber accuracy 0.2% (volume 500 ml ± 10%)

 

  • Tube

 

 

 

 

NBP problem

 

Symptoms Possible cause Action
The NBP cuff will not inflate. The tube or cuff is bent or twisted. Straighten the tube or cuff.
Air leaks from tube or cuff. Replace the cuff and make sure there are no air leaks in the tube.
The NBP module is malfunctioning.

Replace the NBP module.

replace the motherboard

NBP measurements are not displayed. Wrong size cuff or improper cuff placement. Use the correct cuff size and ensure correct cuff placement.
There is a problem external to the monitor. Make sure the tubing is straight and free of kinks. Replace the cuff and make sure there are no air leaks in the tube.
Incorrect initial inflation pressure setting. Ensure all external blood pressure reading requirements are met and the patient is not excessively mobile
An NBP module error has occurred.

Replace the NBP module.

replace the motherboard

NBP measurements are not reliable The cuff is the wrong size, or the cuff is placed incorrectly. Use the correct cuff size and ensure correct cuff placement.
There is an external problem. Ensure that all external blood pressure reading requirements are met and that the patient is not excessively mobile.
Incorrect initial inflation pressure setting. Specify the correct initial inflation pressure.
Calibration is not reliable. Recalibrate the NBP module
NBP initial pressure is too high. The cuffs are too tight. Make sure the cuff is properly fastened properly.
Incorrect initial inflation pressure setting. Specify the correct initial inflation pressure.

 

 

 

Removing the NBP module

 

Caution

 

After replacing the NBP module, the NBP cycle count must be reset and the NBP test run.

 

 

 

1.Separate the front panel from the rear case assembly.

 

2. Disconnect the NBP cable from the motherboard.

 

 

 

3.Disconnect the NBP tubing from the motherboard.

 

 

4.Remove the three Torx T10 screws from the metal bracket that attaches the NBP module assembly to the motherboard

 

 

 

5.Remove the assembly from the front housing.

 

FAQ

 

1.Is the service life of the accessories the same as that of the original parts?

Yes, it is. Our accessories are made from the same materials and meet the same manufacturing standards as the original parts, ensuring a nearly identical service life.

 

2.Are your accessories compatible with portable ultrasound devices?

Yes. We offer specialized accessories designed for portable ultrasound devices, which are compact and lightweight, compatible with various types of portable ultrasound equipment.

 

3.Can your accessories be compatible with obsolete medical equipment?

Yes. We maintain a long‑term inventory of spare parts for outdated medical devices; by providing the equipment model, we can match the corresponding compatible components.

 

4.What information is required for compatibility verification?

The most critical information is the device model and photo of the nameplate; providing photos of accessories can expedite the verification process.

 

5.What should I do if the installation hole position is off?

If the deviation is caused by manufacturing errors, we will replace the parts for you free of charge to ensure proper installation fit.

 

Thank you for your attention .If you're interested in PHiliphh VS2+ Patient Monitor Parameter NBP Spo2 Panel Board Connector 453564200151 ,feel free to contact us .Thank you .

 

 


90 Day Warranty                             SGS ceryification

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.