Mindray IMEC8 Patient Monitor parts 8'' Touch Screen With 4 Wire Line
Mindray IMEC8 Patient Monitor 8” 4 Wire Line Touch Screen
Mindray IMEC8 Patient Monitor 8”Touch Screen With 4 Wire Line
Touchscreen and Touchscreen Control Board
Fully configured iMEC patient monitor supports touch screen.
The touch screen can be controlled and calibrated by touch. The touch screen is connected to the touch screen control panel and the motherboard.
The touch screen control panel controls the touch screen and communicates with the monitor. The touch screen control board communicates with the motherboard through I2C interface and connects to the touch screen through FPC
Touchscreen Calibration
1.Select [Main Menu]→ [Maintenance >>] → [Cal. Touchscreen].
2.The 
3.Select, in turn, the central point of the 
4. After the calibration is completed, the message [Screen Calibration Completed!] is displayed. Select [Ok] to confirm the completion of the calibration.
Troubleshooting
| Symptoms | Possible Cause | Corrective Action |
|
Touchscreen does not respond |
Touchscreen disabled |
Check if there is a symbol
|
| Cables defective or poorly connected |
1. Check that the cables from the touchscreen to the touchscreen control board, the touchscreen control board to the button board, and the button board to the mother board are correctly connected. 2. Check that the cables and connectors are properly connected
|
|
| Touchscreen control board defective | Replace the touchscreen control board | |
| Touchscreen defective | Replace the touchscreen | |
| Main board defective | Replace the main board |
Specification Of Mindray IMEC8 patient monitor 8” touch screen
| Product name |
Mindray IMEC8 patient monitor 8” touch screen
|
| Brand | Mindray |
| Application | Mindray IMEC8 patient monitor |
| Size | 8” |
| Material | Glass |
| MOQ | One piece |
| Condition | Original new |
| Warranty | 90 days |
Removing the Touchscreen Control Board (Optional)
Loose and unscrew the two M3X8 screws as shown below. Disconnect the touchscreen cable and cable for touchscreen control board, and then
remove the touchscreen control board
Removing the Touchscreen (Optional)
Unscrew the ten M3X8 screws indicated below. Take out the touchscreen bracket and then the touchscreen
Disassembling the Screen
1. Unscrew the ten M3X8 screws indicated below. Then take out the screen bracket.
2. Unscrew the four M3X8 screws indicated below to remove the screen