► Brief Introduction
This NIBP Board is working for PHiliphh Goldway UT4000F Patient Monitor. It is responsible for processing and displaying the vital signs and other information of the patient. The main board is a critical part of the monitor and is essential for its proper functioning.
► Non-invasive Blood Pressure Test and Calibration
The accuracy of the NIBP reading of the monitor is decided by: NIBP calibration and the measurement range of the blood pressure sensor. Correct testing and calibration methods are as follows:
- Unplug all cables connected to the monitor except for the power cable.
- Turn on the power switch of the monitor.
- Connect the hose of the cuff to digital pressure meter, auto inflating bladder and NIBP socket at the side panel of the monitor through a 3-way TEE. Inflate the cuff to a certain pressure level, compare the pressure reading on the digital pressure meter and the monitor screen to see if their value difference are less than 3 mmHg. If not, you need to do blood pressure calibration.
- Input a password and enter the system maintain menu as follows
- Rotate knob to Pressure Compensation, select the value to be calibrated (-6 – 6mmHg or Auto), then exit the system calibration menu.
- Repeat (3) , if digital pressure meter and monitor screen have the same pressure value, then calibration is complete. If not, re-calibration is required. If the same results cannot be obtained through repetitive calibration, please contact manufacturer.
► Abnormal NIBP reading
Phenomenon: NIBP measurement result is abnormal.
Inspection method: Check if the cuff leaks or the tube connector leaks.
Solution: Replace it with good cuff and connector.
► Specification
| Method | Automatic oscillometric |
| Parameters | Systolic, diastolic, mean arterial pressure, pulse |
| Scale | mmHg or kPa |
| Operating Modes | Manual, Automatic, Continuous |
| Repeat Cycles | 1 ~ 10, 15, 30, 60, 90, 120 minutes |
| Deflation Pressure | 30mmHg(4.0kPa) higher than the last systolic pressure. |
► Company Business
CWT Technology HK Limited, we are a trustworthy medical equipments supplier as we have been engaged in this industry for many years. During these years of marketing and expanding, we have accumulated a great wealth of experience both in product distribution and customer service. Therefore, please rest at ease, you will receive satisfactory goods if you choose our products. We have been concentrating on producing and providing top quality items and first-class service. We firmly comply with the belief that customers and prestige first. If you want to explore more or have any questions about us, please feel free to contact us.
Thank you for your time and interest in PHiliphh Goldway UT4000F Patient Monitor NIBP Board 4C1N01A.
If you want to explore more details please let us know.
Any inquiry will be highly appreciated and thanks for your support.
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.