10097559 PSR-11-75-KE13PB oxygen sensor for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine

 

 

Oxygen sensor calibration test

 

Test the oxygen sensor calibration as follows:

 

1. Connect the ventilator oxygen tubing to a known source of 100% O2 (e.g., a medical oxygen cylinder). Press the 100% O2/CAL button for 2

 

minutes or the INCREASE O2 button for 2 minutes to calibrate the oxygen sensor. Once the button light turns off, proceed to the next step.

 

2. Connect the ventilator oxygen tubing to another known source of 100% O2 (e.g., a second medical oxygen cylinder).

 

3. Set %O2 to the following values ​​and wait one minute for each monitored value to stabilize: 21% 40% 90%

 

4. Observe the upper display to ensure that the O2 value (% O2 delivered) is within 3% of each setting within one minute of selecting each

 

setting.

 

Oxygen Sensor Calibration Button: Older ventilators use the 100% O2/CAL 2 minutes button, while newer ventilators use the INCREASE O2 2

 

minutes button. Provides 100% oxygen (if available) for two minutes and calibrates the oxygen sensor. The green light on this button will

 

illuminate and a message (100% O2 Cal in Progress) will appear on the lower touchscreen to indicate 100% O2 delivery is active. If you press the

 

O2 button again, the system will restart the two-minute delivery interval. Press CANCEL to stop calibration. For information on calibrating the

 

oxygen sensor, see page TR 15-5.

 

 

Alarm Messages

 

If you see this O2 Sensor message, it means that background checks have identified a problem with the oxygen sensor (sensor error or not calibrated). Patient ventilation will not be affected. And you can:

 

• Press 100% O2 CAL or O2 INCREASE for 2 minutes to recalibrate the oxygen sensor.

 

• Disable the oxygen sensor.

 

• Replace the oxygen sensor.

 

 

 

Specification of 10097559 PSR-11-75-KE13PB oxygen sensor for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine

 

 

Product Name

PSR-11-75-KE13PB 10097559 oxygen sensor for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine

 

 

Brand Covidien
Part No

 

PSR-11-75-KE13PB 10097559

 

Application

PB760 PB840 PB980 Puritan Bennett Ventilator Machine

 

MOQ One piece
Condition Original /Compatible new
Warranty 90 days

 

 

 

Photo 

 

Below photo is PSR-11-75-KE13PB 10097559 oxygen sensor ,which can be used for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine .For more details ,please feel free to contact us .

 

 

Thank you so much for your attention .If you're interested in 10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine ,feel free to contact us .Thank you .

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.