Fukuda CardioMax FX-3010 ECG Machine Mainboard PCB-6380D-C1 Used in Good Working with 90 Days Warranty

Price Negotiable
Price: Negotiable
MOQ: 1
Delivery Time: 5-8 work days
Brand: Fukuda
Product Description
Fukuda CardioMax FX-3010 ECG Machine Mainboard PCB-6380D-C1 Used in Good Working
The Mainboard PCB-6380D-C1 is the core component of the FX-3010 ECG Machine. This mainboard receives microvolt-level bioelectrical signals from the patient's leads. It coordinates and manages all operations of the machine, including keyboard input, LCD display, thermal printhead drive, motor control, and external communication. It is responsible for analog-to-digital conversion, filtering, analysis, and temporary storage of signals, providing the final waveform data for printing and display. Any ECG signal acquisition-processing-output process must pass through the precise processing of this mainboard.
Specification
Brand name Fukuda
Model Number FX-3010
Product Name Mainboard
MOQ 1 Piece
Condition Used original
Warranty 90 Days
Delivery Method DHL, Fedex, UPS
Payment Terms Full Before Shipment
Features Professional, reliable, Economic
Stock Situation Stock available
Packing Carton
Delivery 3-5 Working Days
Method of Transport Expressage
Use for CardioMax FX-3010 ECG
Fukuda CardioMax FX-3010 ECG Machine Mainboard PCB-6380D-C1 Close-up view of Fukuda CardioMax FX-3010 ECG Machine Mainboard PCB-6380D-C1
Main Functions of the Motherboard
  • Preamplifier and Lead Selector: This is the first stage of signal processing. It uses a multiplexer to switch different electrode combinations based on the user-selected lead mode (I, II, III, aVR, aVL, aVF, V1-V6). Subsequently, a high-input-impedance, low-noise instrumentation amplifier provides initial amplification of the weak millivolt-level ECG signal.
  • Analog Signal Conditioning Circuit: Further processes the initially amplified signal. This mainly includes:
    • Filtering: Removing baseline drift (such as that caused by breathing or exercise) using a high-pass filter, and suppressing high-frequency noise such as EMG interference using a low-pass filter. Typically, a powerful power frequency notch filter is also included to specifically filter out 50/60Hz power supply interference.
    • Amplification: The main amplifier amplifies the signal to a level suitable for processing by the analog-to-digital converter (ADC).
  • Analog-to-Digital Converter: Converting the conditioned, clean analog ECG signal into a digital signal for subsequent digital processing and analysis by the microprocessor. This is a crucial step in the transition of signals from the "physical world" to the "digital world."
  • Processor and Digital Signal Processor: This is the "brain" of the motherboard. It is responsible for executing digital filtering algorithms to further optimize the waveform and running ECG analysis programs to perform automatic measurements (such as heart rate and intervals) and arrhythmia analysis.
Thank you for your attention. If you have any questions about Fukuda CardioMax FX-3010 ECG Machine Mainboard PCB-6380D-C1 Used in Good Working, please contact us freely.

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.