Mindray BC-30s Fully Automatic Three-Group Blood Cell Analyzer (Used)
The Mindray BC-30s fully automatic three-group blood cell analyzer is a pre-owned device manufactured by the original manufacturer. Professionally tested and fully functional, this cost-effective solution is ideal for medical facilities seeking reliable hematology analysis.
This analyzer rapidly detects 18 routine blood test parameters including white blood cells, red blood cells, and platelets, generating 3 histograms from whole blood samples. Perfect for primary care hospitals, community health centers, and clinics, it delivers an optimal balance of performance and affordability.
Key Features
- Three-part differential technology: Provides accurate differentiation of neutrophils, lymphocytes, and intermediate cell groups, meeting primary care diagnostic requirements with stable output and minimal maintenance.
- Fully automated operation: Processes samples without manual intervention, handling 150-200 samples daily - ideal for high-volume outpatient testing.
- Durable construction: Built with Mindray's corrosion-resistant components that maintain stability even after extended use, reducing maintenance frequency.
- Advanced data management: Features hospital system connectivity for data upload and includes quality control protocols for reliable results.
Technical Specifications
| Model | Mindray BC-30s |
|---|---|
| Detection Type | Fully Automated Three-Part Differential Blood Cell Analysis |
| Parameters | 18 detection parameters + 3 histograms |
| Sample Volume | 20 μL (whole blood) |
| Testing Speed | ≤60 seconds/sample |
| Daily Capacity | 150-200 samples |
| Display | 10.1-inch color touchscreen |
| Data Storage | 10,000 test results |
| Power Requirements | AC 220V±22V, 50Hz±1Hz |
| Power Consumption | ≤120W (operating), ≤30W (standby) |
Frequently Asked Questions
Q: What preparations are needed before reusing a BC-30s after long inactivity?
A: Verify all hardware components including device appearance, voltage, and reagent tubing. Most importantly, conduct performance verification - the device can only be used if values are within normal range.
Q: What causes persistently low platelet counts and abnormal histograms?
A: First check sample quality (anticoagulant amount and testing timeframe). If normal, clean tubing to remove reagent residue and calibrate the platelet detection channel according to the manual.
Q: How to resolve sudden increase in reagent consumption?
A: Check for tubing leaks first. If none exist, verify "reagent dispensing volume" settings and adjust cleaning frequency to once every 10 samples to conserve reagents.
Ordering Information
Payment Methods: Bank Transfer / PayPal / Western Union
Shipping Methods: DHL / FedEx
All products include a 90-day warranty. Contact our after-sales service for any warranty claims.
Shipping Methods: DHL / FedEx
All products include a 90-day warranty. Contact our after-sales service for any warranty claims.
Related Videos
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.
Company
China World Technology Medical Equipment Service Group
Location
TianHe District,GuangZhou,GuangDong
Contact Person
Kiara