PHiliph UT3000A monitor parameter panel 35201305000 Medical Equipment Parts
PHiliph UT3000A monitor parameter panel 35201305000 Medical Equipment Parts
The PHiliph UT3000A monitor parameter panel, part number 35201305000, is an original, salvaged parameter panel from the PHiliph UT3000A monitor. It is functional and can be used as a repair or replacement part for the UT3000A monitor. The parameter panel integrates indicator lights and interfaces, providing interfaces for sensors and performing signal conversion. It works in conjunction with the navigation knob and mainboard, forming a core component for human-machine interaction.
The parameter panel also features labels with corresponding interface prompts, such as "SpO₂" and "NIBP," allowing medical personnel to easily locate the correct interface and save operation time.
Key Features
1. Signal Conversion
Provides an interface for the sensor, enabling signal conversion and transmission to other components for analysis.
2. Labeling and Warnings
The panel is labeled with the corresponding function of each interface (e.g., "SpO₂", "NIBP") and safety warnings to facilitate correct connection by clinical personnel.
3. Strong Compatibility
As seen in the product details image, the interface board has reserved holes for SpO₂ and NIBP interfaces, ensuring compatibility with the high-end UT3000A monitor.
4. Enhanced Aesthetics
The overall color scheme corresponds to the main body of the monitor, creating a harmonious look and enhancing the overall aesthetics of the device.
|
Product name |
parameter panel |
|
Model |
35201305000 |
|
Machine Model |
UT3000A |
|
Condition |
Tested & Guaranteed |
|
Warranty |
90 days |
|
Material |
Plastic+Metal |
|
Shipping by |
DHL/FedEx/UPS/Etc. |
|
Shipping time |
7-10 days |
|
Payment terms |
Bank TT/ Paypal /Western Union |

Q1: The monitor displays no sensor signal. Is this an interface board malfunction?
A1: Connect the sensor to another device to check for sensor malfunction. If the sensor is normal, check if the communication between the parameter panel and the main unit is interrupted, and check the wiring connections. If communication is normal, replace the parameter panel.
Q2: The monitor keeps interrupting data acquisition due to a loose interface. How can this be resolved?
A2: Check if the interface pins are damaged, such as deformed or bent pins. Replace the interface board promptly to avoid affecting monitoring efficiency.
Q3: Can I replace the interface board myself?
A3: The interface board is secured with clips and screws, making replacement relatively simple. You can replace it yourself, but be careful not to connect sensors during replacement to avoid damaging them.