Mindray Datascope Passport 2 Patient Monitor With Rear Cover and Handle TM 08517-L3

 

Product Introduction:

 

This is the original factory-installed back cover and handle assembly for the Mindray Datascope Passport 2 patient monitor, part number TM 08517-L33. Made of white medical-grade engineering plastic, it consists of a back cover and a handle, allowing medical staff to easily lift and quickly transport the equipment. It is ideal for scenarios involving bedside movement, emergency resuscitation, and intra-hospital transport. The design and size are completely consistent with the main monitor unit, making it an important accessory for protecting the rear of the monitor.

 

 

Rear Panel:

 

45.AC Receptacle
Insert an AC power cord into this connector.
CAUTlON: Use only Mindray DS supplied power cords, or if a substitute isnecessary, use only hospital grade power cords.


46. Equipotential lug

Provides Equipotential grounding of hospital equipment


47. Soft Grip Handle
Use for carrying the monitor.


48.Main I/O Connector Port (DM1)

Area dedicated for the use of an optional communication port.


49.Expansion Slot

Used for connecting an optional Panorama Instrument Radio .608.

 

Main Functions:

 

1. Rear Cover Protection: Fits snugly against the rear of the monitor unit, blocking dust, liquids, and minor impacts, protecting internal circuit boards, interfaces, and other components from damage caused by the daily clinical environment. It also assists the main unit's ventilation holes in maintaining airflow.

 

2. Portable Handle: The integrated handle is ergonomically designed, facilitating rapid equipment transport within the hospital by medical staff (from the emergency room to the ward, or from the ward to the resuscitation room), without the need for additional carts or other tools.

 

3. Structural Reinforcement: The snap-fit design of the rear cover fits seamlessly with the main unit, and the handle's bracket is integrally molded with the rear cover, enhancing the overall structural stability of the equipment and reducing swaying during movement.

 

Key Advantages and Features:

 

1. Precise Factory Fit: The snap-fit positions on the back cover and the mounting holes for the handle are custom-molded for the Passport 2 requiring no additional processing or adjustment. It fits perfectly to the main unit without any looseness or misalignment.

 

2. Durable and Practical: The outer shell is made of anti-aging medical-grade plastic, scratch-resistant and impact-resistant. The handle's load-bearing design meets the needs of frequent clinical movement and is not easily deformed or broken even after long-term use.

 

3. Integrated Design: The back cover and handle are matching components, resulting in a unified appearance without being obtrusive. The handle's reasonable placement ensures a stable center of gravity when carrying, reducing the burden on medical staff during transport.

 

 

Disassembly Guide:

 

Before installing or removing the Rear Cover , perform the following:

 

• Power down the Mindray Datascope Passport 2 Monitor and remove the AC line cord.

• Disconnect all cable assemblies from the left side, right side and rear of the monitor.

• Remove any installed batteries from the monitor.

• Perform all work on a properly grounded work station and place a soft cloth on the surface to protect the monitor.

 

To Removal of the Rear Cover and Handle Assembly:

 

1. Lay the monitor face down on the soft protective surface to avoid scratching the front screen and keypad.

2. Locate the fixing screws (or snaps) on the handle bracket. Use a suitable screwdriver to loosen and remove the 2 screws ; if using snaps, gently pry open the handle snaps with a plastic tool.

3. Lift the handle assembly vertically upward to separate it from the rear cover.

4. Remove the 6-8 fixing screws from the edge of the rear cover (positions correspond to the monitor’s rear housing). For snap-on rear covers, use a plastic pry tool to gently separate the snaps along the cover edge.

5. Grasp the rear cover with both hands and pull it backward gently to separate it from the monitor main unit.

6. If there is a ground strap or auxiliary connection wire between the rear cover and the main unit, carefully disconnect it from the corresponding connector.

 

Usage Precautions:

 

•Cleaning Method

Wipe with a soft, slightly damp cloth, or a small amount of neutral detergent (such as diluted soapy water). Never use abrasives, alcohol, or strong corrosive cleaners, as these may damage the casing.

When cleaning, prevent any liquid from flowing into the gaps, interfaces, or vents of the device casing. Never immerse or rinse the device.

•Handle Use

When moving the device, always use its dedicated handle for lifting. Do not pull on the cables or other parts of the monitor; the handle is designed to bear weight.

•Avoid Physical Damage

Prevent bumps, drops, or heavy pressure during use, especially on the screen and corners. When not in use, place it on a stable surface or a dedicated trolley.

•Environment and Storage

Store the device in a dry, clean, and well-ventilated environment, away from high temperatures, heat sources, and direct sunlight to prevent aging or deformation of the casing.

 

FAQ

1. Could you provide your company's brochure or promotional materials?

Yes, we will offer both Chinese and English versions of the company brochure and product manuals.

 

2. What is the inventory status of accessories? Are popular accessories in stock?

Inventory levels fluctuate frequently; we will reserve stock for you after you place an order.

 

3. Can you provide customized medical accessories?

Yes, we offer OEM services and can deliver professional customization.

 

4. Does the accessory quotation include tax?

The quotation is generally FOB (Free On Board) price excluding tax; if tax inclusion is required, it must be explicitly stated.

 

5. Do you possess proprietary R&D patents?

You hold over ten utility model and design patents.


90 Day Warranty                             SGS ceryification

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.