M3015-66442 PHiliph M3015A Module Main Interface Control Circuit Board Replacement Parts


  • Product Description

 

  • This is the main board inside the M3015A Microstream CO2 expansion module. It serves not only as an interface board but also as the core control board for the module.

 

  • One side of the board features a dense row of pins, which serve as the interface for connecting to the M3001A host or the backplane of the MP40/50 monitor.

 

  • It communicates with the host via the MSL (Measurement Server Link) protocol for high-speed data exchange, transmitting collected CO2 data to the host for display while receiving control commands from the host.

 

  • This green printed circuit board (PCB) integrates high-density surface-mount components, a central processing unit, and a rugged edge connector to ensure seamless integration.

 

  • Core Functions:

 

  • Its primary function is to serve as the “brain” of the carbon dioxide monitoring system, processing signals from the microfluidic CO2 sensor and transmitting critical patient data to the host via the Measurement Server Link (MSL) interface.

 

 

 

 

  • Product Specifications

 

  • Part Number: M3015-66442
  • Component Type: Main Interface Board
  • Compatible Module: PHiliph M3015A
  • Compatible Hosts: Intellivue MP40, MP50, M3001A
  • Form Factor: Single-width module PCB
  • Connectivity: High-density edge connector
  • Application: Medical patient monitoring

 

Packing Standard Export Carton
Origin Germany
Operating Temperature

0°C to 40°C

Storage Humidity <60% RH
Certification Standard IEC 6060

 

  • Common Failure Points

 

1. Connector Wear: Discoloration or wear on the gold pins at the bottom may cause the module to fail recognition or experience communication interruptions.

 

2. Power Management Chip: Located within the three square silver shields, this component controls power supply. If the module fails to power on, the issue typically originates here.

 

3. Main Processor: The large chip on the left. Damaged firmware or hardware failure in this component causes the module to fail self-test (displaying a red exclamation mark or “Module Error”).

 

  • Repair and Replacement Recommendations

 

1. Disassembly Difficulty: This component is located deep within the equipment. Replacing it requires first removing the bottom cover, pump, and CO₂ absorber before finally detaching this interface board.

 

2. Safety Warning: High voltage (800V) is present inside the equipment, and internal components may be contaminated with bacteria. Only qualified maintenance personnel may perform this operation after disconnecting power and wearing protective gloves.

 

FAQ

 

           1.Can universal accessories be used interchangeably among different brands?

           Some universal accessories of the same type can be used interchangeably; please provide device model for verification.

 

           2.What should I do if minor process variations affect usage?

           We guarantee all batch‑produced accessories meet original standards; free replacement if functional issues occur.

 

           3.Is the insertion‑extraction force of the accessory interface consistent with the original?

           Yes, precisely calibrated to ensure secure connection and easy operation without damage.

 

           4.Can we provide on‑site compatibility testing services?

           Yes, for bulk orders we provide on‑site testing and demonstration services.

 

           5.What should I do if the compatibility verification result is inconsistent with actual use?

           We will re‑test free of charge and provide corresponding solutions or replacement services.

 


90 Day Warranty                             SGS ceryification

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.