This product is a multi-parameter board assembly specifically designed for the Q5 patient monitor, integrating the complete NIBP (Non-Invasive Blood Pressure) measurement module along with supporting circuitry for other physiological parameters. The board features a precision-engineered PCB with dedicated sections for pneumatic control, pressure sensing, and signal processing. Its layout includes high-quality electronic components such as pressure transducers, solenoid valves, and analog-to-digital converters, all arranged for optimal signal integrity. The NIBP section operates on the oscillometric measurement principle, controlling cuff inflation and deflation while analyzing pressure oscillations to determine systolic, diastolic, and mean arterial pressure values. This parameter board serves as the measurement control center for vital signs acquisition, supporting adult, pediatric, and neonatal patient populations with clinically validated accuracy.
Biolight Q5 parameters
Including ECG(3-lead or 5-lead or 12-lead selectable), arrhythmia detection,ST segmentanalysis, Heart Rate (HR), Respiration Rate (RESP), Temperature (TEMP),Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure(NIBP), Invasive Blood Pressure (IBP),Carbon dioxide (CO2), Anesthetic GasAG)
Parameter Specifications:
ECG
• Leads: I, II, III, aVR, aVL, aVF, V-
• Mode: 3-leads, 5-leads
• Gains: 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, 40mm/mV, auto
• Scanning Speed: 12.5mm/s, 25mm/s, 50mm/s
• Heart Rate Range: Adult: 10~300bpm Child &Neonate: 10~350bpm
•Accuracy: ±1bpm or ±1% whichever is greater
NIBP
• Operation Modes: Automatic Oscillometry
• Range: Adult: 10 ~270mmHg Child: 10 ~235mmHg Neonate: 10 ~135mmHg
• Measurement: Systolic Pressure/Diastolic Pressure/Mean Pressure
• Unit: mmHg, kPa
• Pulse rate range: 40 ~240bpm
SpO2 (Pulse Oximetry)
• Range: 0 ~100%
• Accuracy: ±2% (70% ~100%); Unspecified (0 ~69%)
• Range & Accuracy of Pulse Rate: 25 ~254bpm ±1bpm or ±1% whichever is greater
Parameter Alarm
You can configure parameter alarms in the dedicated settings menu for each parameter. Within the menu of a specific parameter, you can view and adjust the alarm limits and alarm status, and the alarm settings for different parameters are independent of each other. You can also set parameter alarms by pressing the 【Main Menu】 smartkey and selecting 【Alm Setup】.
For parameters with the alarm switch set to ON, an alarm will be triggered once the parameter value exceeds the preset alarm limit. When an alarm occurs, the following actions will be executed:
-
Audible and visual alarms will be activated according to the alarm level.
-
The corresponding alarm message will be displayed in the physiological alarm area.
-
The value of the parameter in alarm will flash continuously.
-
If alarm recording is enabled, the recorder will start recording alarm data at the preset interval.