Hospital Medical Equipment Parts PHiliphhilip MP70 Patient Monitor ECG Out Board M8085-67001 CPU Board

 

 

 

► Brief Introduction

ECG-Out Marker-In

ECG Out/Marker In interface Receives ECG waveforms directly from the ECG/Resp arrhythmia ST-Segment PHiliphhilipysiological algorithm through the RS-422 serial interface and converts digital ECG signals to analog ECG signals. In addition, the ECG Out controller receives labeling information from the connected device and forwards this data to the ECG/Resp arrhythmia ST segment PHiliphhilipysiological algorithm. The converted analog signal is used to synchronize the connected device with the patient's electrocardiogram

RS-422

The SERIAL link RS-422 interface transmits the ECG signal to the ECG output/label input of the IntelliVue patient monitoring system display unit. The interface is a serial, differential, full-duplex link. The interface has ESD protection.

PS/2

The PS/2 interface supports the serial protocol for standard PS/2 devices (mice). PS/2 serial protocol is interpreted by the HIF control interface.

 

 

Removing the ECG Out board

For instructions on removing the ECG Out board when an SRL2 board is plugged

1 Remove the mounting plate from the monitor if attached.

2 Remove the four screws with a T20 screwdriver and pull off the bottom cover.

3 Unscrew the screws on the ECG Out board and pull on the tab to remove the cable.

 

 

Troubleshooting the ECG OUT

Cause of Failure Failure Isolation and Remedy
No marker pulse is
displayed on the Monitor
or no ECG- OUT signal to
the Defib
Disconnect the MMS and Defib cable.
Switch the Monitor off then on again.
Observe the red LED in the ECG OUT section. (Note that the LED can only be observed if the housing bottom is removed).
ECG OUT Board defective Replace ECG OUT board
ECG OUT Board defective If the red LED switches on and remains on for more than 20 seconds:
Replace ECG OUT Board
ECG OUT Board defective Connect Known good Defib, Defib cable MMS and MMS cable. Check Marker pulse and ECG OUT signal at defib again. If there is still no signal:
Replace ECG OUT board

 

 

ECG Out Functionality

Connections

The cable M1181A #A62 has both ends terminated. The PHiliphhilipotograPHiliphhilip above shows the monitor side

connection.

If using a non-terminated cable:

  1. Strip 5 mm (3/16") insulation from leads and twist conductor strands tightly.
  2. Solder leads to the connector as shown in the following diagram.

 

 

► Delivery Clauses

To make sure the work goes smoothly, please pay special attention to him and arrange your time to apply carefully according to your personal needs.For delivery time, we will do our best to process and ship orders as soon as possible. Please be patient and stay in touch with us. In general, most orders will be shipped within 3 to 5 business days and can arrive on time. However, during busy days, the shipment may be delayed by more than 5 business days. If you are in dire need of your product, please be sure to confirm with us before submitting your applications. If your order takes longer than expected or if you want to clarify that upon arrival, feel free to contact us. We're always here to help you.

 

 

Thank you for your time and interest. If you have any questions about PHiliphhilip MP70 Patient Monitor ECG Out Board M8085-67001, please feel free to contact us at any time, we are glad to answer them. Customer satisfaction. Our biggest pursuit!

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.