Product Description

OOM102-1 OOM102 OOM202 OOM201 Envitec Oxygen Cell Sensor 100% Genuine New

 

The EnviteC medical oxygen sensor is designed to be used as the oxygen sensor component of the oxygen analyzer to measure the oxygen concentration in the following breathing gas mixture

application:

Sensing device for oxygen in

• Oxygen concentrator control device

• Medical ventilator

• Anesthesia equipment

• Incubator.

Use is limited to system monitoring. The sensor is not suitable for breath analysis of breathing gas. Please refer to

To the instructions for use! If the sensor is going to be replaced

The original oxygen sensing component of the oxygen analyzer,

Please refer to the EnviteC XRL cross-reference list to select

The right sensor

 

 

Technical Specifications OOM202

 

Model Number OOM202
Nominal Sensor Lifetime 250000 Percent Volume Oxygen Hours
Electrical Interface Gold Plated Slip Rings
Output Signal in Air 170A - 230A No Temperature Compensation
Response Time T 90% Less Than 12 Seconds
Weight Approx. 28 Grams
Usage Hospital

 

 

Technical Specifications OOM102-1  OOM102

 

Model / Part Number OOM102-1
Target gases OXYGEN
Measuring Range 0-100%
Operating Voltage 5V
Oxygen Pressure Range 2 bar

 

Technical Specifications OOM201

 

Measurement range 0 % ... 100 % oxygen (at atmosPHiliphhiliperic pressure)
Nominal sensor lifetime ≥ 500 000 % volume oxygen hours
Output in ambient air 14 mV ... 20.7 mV (Dual Cathode), load 600 Ohms
Electrical interface Gold plated slip rings
Accuracy meets ISO 80601-2-55 requirements
Repeatability < 1 % volume O2 at constant temperature and pressure
Linearity error < 3 % relative
Response time < 12 s to 90 % of final value
Zero offset voltage < 200 µV in 100 % nitrogen, applied for 5 min
Cross interference meets ISO 80601-2-55 requirements
Influence of humidity -0.03 % rel. per % RH at 25 °C
Pressure range 0.6 bar ... 2 bar (ppO2 0 ... 1250 mbar O2 )
Influence of pressure proportional to change in oxygen partial pressure
Influence of mechanical shock < 1 % relative after a fall from 1 m
Operating temperature 0 °C ... +50 °C
Temperature compensation no temperature compensation
Operating humidity 0 % ... 99 % RH non-condensing
Long term output drift < 1 % volume oxygen per month typically < -15 % relative over lifetime
Storage temperature -20 °C ... +50 °C
Recommended storage +5 °C … +15 °C
Recommended load ≥ 10 kOhms
Warm-up time < 30 minutes, after replacement of sensor
Weight approximately 28 grams
Part number 01-00-0014

 

Mechanical drawing (All dimension in mm)