Ventilator Sterilizable Exhalation Expiration Valve Disinfected Assembly 115-021461-00 Mindray SV300 SV600 SV800 E3 E5
Mindray SV300 SV600 SV800 E3 E5 ventilator sterilizable exhalation Expiration Valve Disinfected assembly EXP valve PN 115-021461-00
► Brief Introduction
The Sterilizable expiration valve assembly is mainly responsible for functions such as pressure control, pressure monitoring and flow monitoring of the patient’s expiratory PHiliphhilipase. Thus, the various components of the expiration module must be cleaned and disinfected before being reused.
The patient’s exhaled warm and moist gas is condensed when it flows through the expiratory hose. The condensed water remains on the hose wall and finally enters the water trap. Whenthe patient’s exhaled gas arrives at the expiration valve, condensed water may appear at the expiration valve (including the expiratory flow sensor), compromising the measurement accuracy of expiratory flow sensor.
Check the Sterilizable expiration valve assembly for water build-up when abnormal flow waveform or unstable tidal volume fluctuation is detected. If there is water build-up inside the expiration valve, clear it before use.
Check the expiratory water trap for water during the use of the ventilator. If there is water build-up, empty it promptly. Water condensation in the expiration valve can be reduced by using a bacteria filter between the expiratory tube and expiration valve
► Expiration Valve Assembly and Membrane

B: Expiration valve handwheel
C: Expiration valve membrane
D: Locked state of the expiration valve
E: Unlocked state of the expiration valve
► To disassemble:
- Rotate the expiration valve handwheel until the indicating arrow
position. Then pull the expiration valve assembly out of the assembly horizontally.
- Remove the expiration valve membrane.
► To install:
- Install the expiration valve membrane onto the expiration valve assembly.
- Ensure that the indicating arrow position.
► Methods for Cleaning and Disinfection
| Parts | Recommended frequency Interval | Cleaning | Disinfection |
| Expiration valve membrane (silicone) | Each patient/weekly | Immersion | Immersion or Autoclaving |
| Expiration valve assembly (except membrane) | Each patient/weekly | Immersion | Immersion or Autoclaving |
- Immersion: Flush with water first and then immerse it in alkalescent detergent (soapy water, etc., water temperature of 40ºC recommended) for approximately three minutes. Finally, clean with water and dry completely.
- Autoclaving : Steam autoclave at 134ºC for 10 to 20 minutes (recommended time).
►Delivery Policy
We are shipped the goods worldwide as soon as payment is received, insured and tracked as standard using trusted couriers, such as DHL, FedEx, UPS, etc,.
Getting your equipment to you swiftly and safely is of utmost importance to us. Costs will depend on your location, how quickly you’d like your package, and what you have ordered.
If your order is taking longer than expected or you want to find out when it will arrive, please contact us.
Please kindly contact us, if you are interested in Mindray SV300 SV600 SV800 E3 E5 ventilator sterilizable exhalation Expiration Valve Disinfected assembly EXP valve PN 115-021461-00, Thank you!
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.


position. Then pull the expiration valve assembly out of the assembly horizontally.